Skip to the content

FDA Medical Device Recalls (openFDA)

Stryker Howmedica Osteonics Corp.: GAM Kirschner Wire, Sterile R, T2 K-Wire, Sterile R and T2 K-Wire Recon, Stryker Trauma GmbH, Prof.-Kuntscher-Strasse 1-5, Schonkirchen, Germany 24232; T2 K-Wire, Sterile R Intended for the temporary stabilization of bone segments or fragments

Recall number
Z-0880-2017
Recalling firm
Stryker Howmedica Osteonics Corp.
Product
GAM Kirschner Wire, Sterile R, T2 K-Wire, Sterile R and T2 K-Wire Recon, Stryker Trauma GmbH, Prof.-Kuntscher-Strasse 1-5, Schonkirchen, Germany 24232; T2 K-Wire, Sterile R Intended for the temporary stabilization of bone segments or fragments
Status
Terminated
Initiated
2016-11-09
Posted
not on file
Terminated
2019-03-27
Reason
A review of packaging revealed the seal integrity of the pouch may be compromised. More specifically, there is a potential that the sterile pouch is not sealed at one end due to a manufacturing error.
Root cause
Packaging process control
Action
Stryker sent an Urgent Medical Device Removal letter dated November 9, 2016, to all affected customers via UPS (with return receipt). Stericycle will be handling the returns. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to inform users of the Medical Device Recall and forward the notice to all those individuals who need to be aware within their organization. Return all affected products available at their location to: Stryker C/O Stericycle 2670 Executive Dr., Suite A Indianapolis, IN 46241 Attn: RA 2016-169 - Event XXX Customers were also instructed to complete and sign the enclosed Business Reply Form and return to Stericycle. Customers with questions should call +1-201-851-6683. For questions regarding this recall call 201-831-5000.
Quantity
2,084 units
Distribution
Worldwide Distribution - US (nationwide) Internationally to AU, CA, CL, CH, ES, FR, GB, SE, JP, FR, ZA, and NL.
Lot, serial or model codes
GAM Kirschner Wire - Manufacturing Part Number 12106450S, Lot #'s K0800BB, K0800BC, K0800BD, K0800BE, K0800BF, K0800C0, K081720, K081721, K0810872, K081723, K081727, K082C8B, K084F89, K084F98, K08683D, K08683E, K086841, K08820B, K089AF5, K08E1E1, K0911F4, K09379B, K096A26, K096A2A, K09A2C, K098213, K098215, K099AA0, K09AD33, K09BA4F, K09D564, K09F026, K0A1EF8, K0A1EFA, KoA1EFB, K0A38FB, K0A63AB, K0A63AC and K0A7BC1  T2 K-Wire - Manufacturing Part Number 18060050S, Lot #'s K084FBF, K0920A4, K0A1EFD, K093747, K0937C4, K09BA53, K09BA54, K09BA55 and K09BA56  T2 K-Wire Recon - Manufacturing Part Number 18063030S, Lot #'s K086847, K08E1E9 and K099AA8
510(k) numbers
not on file
PMA numbers
not on file
Product code
NDE
Firm FEI
3003070421
Firm city
Mahwah
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register