Medacta Usa Inc: Versafitcup Impacting Ring Model Number: 01.26.10.0133 (46mm), 01.26.10.0134 (48mm), 01.26.10.0135 (50mm), 01.26.10.0136 (52mm), 01.26.10.0137 (54mm), 01.26.10.0138 (56mm), 01.26.10.0139 (58mm), 01.26.10.0140 (60mm), 01.26.10.0141 (62mm), 01.26.10.0142 (64mm). Designed for cementless use in total hip arthroplasty.
- Recall number
- Z-0878-2013
- Recalling firm
- Medacta Usa Inc
- Product
- Versafitcup Impacting Ring Model Number: 01.26.10.0133 (46mm), 01.26.10.0134 (48mm), 01.26.10.0135 (50mm), 01.26.10.0136 (52mm), 01.26.10.0137 (54mm), 01.26.10.0138 (56mm), 01.26.10.0139 (58mm), 01.26.10.0140 (60mm), 01.26.10.0141 (62mm), 01.26.10.0142 (64mm). Designed for cementless use in total hip arthroplasty.
- Status
- Terminated
- Initiated
- 2010-01-15
- Posted
- 2013-02-23
- Terminated
- 2013-02-23
- Reason
- Medacta USA is recalling Versafitcup Impacting Ring because the impaction ring may not fit properly in the implant shell in all cases. This could result in surgical delay potentially increasing possible complications under anesthesia, having to ream up to a larger shell size leaving less bone than planned and repositiong/removal of the shell after initial impaction.
- Root cause
- Process control
- Action
- Medacta USA sent notification letters dated 1/15/10 to Sales Representatives for customers that purchased the Versafitcup Impacting Ring. The letter informed the customers of the problem identified and the actions to be taken. Customers are informed that Medacta USA personnel will contact them by telephone to provide them with a RMA number. Customers are informed that they will receive a new set of redesigned/reworked impacting rings and a second near the end of the January 2010. Customers with questions are instructed to contact the firm at (805) 437-7085.
- Quantity
- 293 units
- Distribution
- Nationwide Distribution including the states of ID, CO, TX, WI, NY, CT, WA, CA, UT, and AZ.
- Lot, serial or model codes
- Reference Number Lot Number Reference Number Lot Number 01.26.10.0133: 095920, 096050, 096050A; 01.26.10.0138: 095925, 096055, 096055A; 01.26.10.0134: 095921, 096051, 096051A; 01.26.10.0139: 095926, 096056, 096056A; 01.26.10.0135: 095922, 096052, 096052A; 01.26.10.0140: 095927, 096057, 096057A; 01.26.10.0136: 095923, 096053, 096053A; 01.26.10.0141: 095928, 096058, 096058A; 01.26.10.0137: 095924, 096054, 096054A; 01.26.10.0142: 095929, 096059, 096059A.
- 510(k) numbers
- ["K083116"]
- PMA numbers
- not on file
- Product code
- MEH
- Firm FEI
- 3006639916
- Firm city
- Camarillo
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28