PEROUSE MEDICAL: Bard Caliber (CL3030) Inflation Device Product Usage: Inflation Device intended for use during cardiovascular to create and maintain, monitor pressure in the balloon catheter - balloon inflation device for angioplasty procedures

Recall number
Z-0876-2015
Recalling firm
PEROUSE MEDICAL
Product
Bard Caliber (CL3030) Inflation Device Product Usage: Inflation Device intended for use during cardiovascular to create and maintain, monitor pressure in the balloon catheter - balloon inflation device for angioplasty procedures
Status
Terminated
Initiated
2014-10-06
Posted
2014-12-29
Terminated
2017-03-24
Reason
Additional instructions for use due to inability to raise the pressure above 10 atm during inflation phase may result in an incomplete inflation or delay in procedure
Root cause
Device Design
Action
Perouse Medical notified Distributors by Safety Alert Letter sent on October 06, 2014 and and October 07, 2014 by e-mail. The letter identified the defective products and lot numbers and provided a description of the malfunction, risk and mitigation statements. US distributors have been requested to provide the required information to their customers. In addition, the firm has updated device instructions for use. The firm is requiring its customer to fill out a safety notice acknowledgement of receipt form.
Quantity
24821 units
Distribution
US Nationwide Distribution in the states of AZ and CO.
Lot, serial or model codes
All lot numbers beginning with Numbers: 1403,1404,1405,1406,1407
510(k) numbers
["K042449"]
PMA numbers
not on file
Product code
MAV
Firm FEI
3006942524
Firm city
Ivry Le Temple
Firm state
not on file
Firm country
France

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register