FDA Medical Device Recalls (openFDA)

Smith & Nephew Inc.: ***REF 66800694***Qty 10***smith&nephew***RENASYS Port Foam Dressing Kit Port***STERILE***Rx Only.***Made in Mexico for/Fabrique au Mexique pour: Smith & Nephew, Inc., St. Petersburg, FL 33716***Smith & Nephew Medical Limited Hull HU3 2BN England***Lot 123456. Product Usage: The product is a suction device used as an accessory to the foam dressing kits, which is used in conjunction with Smith & Nephew Negative Pressure Wound Therapy (NPWT) systems to deliver negative pressure to a wound.

Recall number
Z-0875-2012
Recalling firm
Smith & Nephew Inc.
Product
***REF 66800694***Qty 10***smith&nephew***RENASYS Port Foam Dressing Kit Port***STERILE***Rx Only.***Made in Mexico for/Fabrique au Mexique pour: Smith & Nephew, Inc., St. Petersburg, FL 33716***Smith & Nephew Medical Limited Hull HU3 2BN England***Lot 123456. Product Usage: The product is a suction device used as an accessory to the foam dressing kits, which is used in conjunction with Smith & Nephew Negative Pressure Wound Therapy (NPWT) systems to deliver negative pressure to a wound.
Status
Terminated
Initiated
2010-12-14
Posted
2012-01-26
Terminated
2012-04-30
Reason
On 12/14/2010, Smith & Nephew, Inc. Wound Management Division in St. Petersburg, FL recalled their RENASYS PORT product code number 66800694 due to small holes in the primary pouch of some of the packaged units of which constituted a breach of the sterile barrier.
Root cause
Nonconforming Material/Component
Action
Smith & Nephew, Inc., Wound Management sent an URGENT DEVICE RECALL letter dated December 14, 2010, to all affected customers. The letter identified the product, problem, and actions to be taken. Customers were instructed to immediately identify and isolate the recalled lots in their inventory, complete and fax the response form, and return any affected products to APL Logistics, Inc. in Suwanee, GA. For questions in general, please contact the Customer Care Center at 1-800-876-1261.
Quantity
71329 units
Distribution
Worldwide Distribution - USA (nationwide) including the states of: AZ, CA, GA, MA, MI, MO, MS,.NC, NY,OR, TN, and WI. and the countries of: Australia, Canada, China, England, Germany, Ireland and Singapore.
Lot, serial or model codes
Product Code Number: 66800694 Lot numbers: 50512962, 50512969, 50512974, 50513087, 50514672, 50513073, 50513074, 50517647, 50516110, 50539644, 50539645, 50539646, 50539647, 50555348, 50555349, 50555351, 50539648, 50546339, 50546340, 50548244, 50548245, 50548246, 50548247 and 50548248.
510(k) numbers
["K082211"]
PMA numbers
not on file
Product code
OMP
Firm FEI
3006760724
Firm city
Saint Petersburg
Firm state
FL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register