Skip to the content

FDA Medical Device Recalls (openFDA)

Micro Therapeutics Inc, dba ev3 Neurovascular: Axium Detachable Coil System QC-7-30-Helix The ev3 Axium Detachable Coil System is indicated for the endovascular embolization of intracranial aneurysms and other neurovascular abnormalities, such as arteriovenous malformations and arteriovenous fistulae.

Recall number
Z-0874-2012
Recalling firm
Micro Therapeutics Inc, dba ev3 Neurovascular
Product
Axium Detachable Coil System QC-7-30-Helix The ev3 Axium Detachable Coil System is indicated for the endovascular embolization of intracranial aneurysms and other neurovascular abnormalities, such as arteriovenous malformations and arteriovenous fistulae.
Status
Terminated
Initiated
2010-03-01
Posted
2012-01-26
Terminated
2012-01-31
Reason
ev3 has initiated a voluntary removal of two lots of Axium Detachable Coil System for product mislabeling as the lots were swapped in production.
Root cause
Employee error
Action
Micro Therapeutics Inc, dba ev3 Neurovascular performed product recovery according to the following outline. a) Quality Assurance Department (QAD) was used to provide details of product distribution for the affected lots. b) Based on the Health Hazard Assessment, the risk was considered undesirable. c) The affected product was part of a limited market release for Physician Preference Testing. As the product was within control of ev3 employees, no public notification was performed. d) Quality Systems contacted the sales representatives identifying the part number, lot number and number of units sent, and a request to segregate and return the affected product was initiated. e) Product returned to ev3 Neurovascular was held in Restricted Stock pending complete device accountability and disposition. f) The ev3 Neurovascular Quality Assurance department scrapped the affected product. g) The ev3 Neurovascular Regulatory Affairs department was responsible for assuring documentation and appropriate notification to regulatory bodies as required.
Quantity
14 units
Distribution
Worldwide Distribution - USA (nationwide) and the countries of the UK, and France.
Lot, serial or model codes
Lot # 7964590
510(k) numbers
["K081465"]
PMA numbers
not on file
Product code
KRD
Firm FEI
3004904811
Firm city
Irvine
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register