MAQUET Inc.: SERVO Ventilator 300A (SV300A); Maquet Critical Care Ab, Solna, Sweden. The device is intended for general and critical ventilatory care of adult, pediatric, infant and neonatal patients. The unit is designed to be used at the bedside and for intra-hospital transport. It is not intended for transport use in ambulances or in helicopters in the US market.

Recall number
Z-0872-2009
Recalling firm
MAQUET Inc.
Product
SERVO Ventilator 300A (SV300A); Maquet Critical Care Ab, Solna, Sweden. The device is intended for general and critical ventilatory care of adult, pediatric, infant and neonatal patients. The unit is designed to be used at the bedside and for intra-hospital transport. It is not intended for transport use in ambulances or in helicopters in the US market.
Status
Terminated
Initiated
2008-12-19
Posted
2009-02-03
Terminated
2009-07-08
Reason
SERVO Ventilator 300A could cause hypoventilation if the use of the PC or PRVC mode with the Automode function set to "On" in Neonatal and Pediatric patient ranges with small tidal volumes and short set inspiratory rise time.
Root cause
Device Design
Action
URGENT: Device Correction letters were sent out on December 22, 2008. Letters alert users to the potential problem and the situation in which it may occur. Letter indicate the set inspiratory rise time should be set in the range of 7-10% and that the humidifier chamber water level be maintained according to manufacturers instructions. The letter is enclosed as a supplement to the SV300A Operating Manual. Consignees were directed to complete a "Device Correction-Customer Fax Response" to indicate number of units of the affected product and that they followed the instructions for including the attached supplemental instructions to the operating manual. For questions about the recall, contact your local Maquet, Inc. representative.
Quantity
3,688 units in US
Distribution
Nationwide
Lot, serial or model codes
All units of SV300A.
510(k) numbers
["K970839"]
PMA numbers
not on file
Product code
CBK
Firm FEI
3008355164
Firm city
Bridgewater
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register