Skip to the content

FDA Medical Device Recalls (openFDA)

Zimmer Biomet, Inc.: ASI 2.0 Calcar Trimmer Shaft for Use with Zimmer Rasp Mode Number: 110032332 Product Usage: The device is a general orthopedic manual surgical instrument

Recall number
Z-0871-2020
Recalling firm
Zimmer Biomet, Inc.
Product
ASI 2.0 Calcar Trimmer Shaft for Use with Zimmer Rasp Mode Number: 110032332 Product Usage: The device is a general orthopedic manual surgical instrument
Status
Terminated
Initiated
2019-12-19
Posted
not on file
Terminated
2020-08-11
Reason
Potential issue associated with the instrument -end of the shaft could fail to effectively mate with the broach
Root cause
Process control
Action
Zimmer Biomet issued letter via email and FedEx on 12/17/19 stating reason for recall, health risk and action to take: Review this notification and ensure that affected personnel are aware of the contents. 2. If you have affected product at your facility, assist your Zimmer Biomet sales representative and quarantine all affected product. Your Zimmer Biomet sales representative will remove the affected product from your facility. 3. Complete Attachment 1 Certificate of Acknowledgement and send to CorporateQuality.PostMarket@zimmerbiomet.com. Questions or concerns after reviewing this notice, please call customer service at 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday.
Quantity
67 units
Distribution
US Nationwide and Foreign: countries of Canada, CHINA, NETHERLANDS
Lot, serial or model codes
Lot Numbers: 827310, 711160, 203240
510(k) numbers
not on file
PMA numbers
not on file
Product code
LXH
Firm FEI
1825034
Firm city
Warsaw
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register