FDA Medical Device Recalls (openFDA)
Remel, Inc: ERIC (Electronic RapID Compendium) Version 1.0.77, Catalog Number R8323600, used in conjunction with: remel RapID Yeast Plus System is a qualitative micromethod employing conventional and chromogenic substrates for the identification of medically important yeast, yeast-like, and related organisms isolated from human clinical specimens. A complete listing of the organisms addressed by the RapID Yeast Plus System is given in the RapID Yeast Plus Differential Chart. RapID" YEAST PLUS System, Catalog No. R8311007. 4 hour identification of yeast based on enzyme technology. Database includes over 40 taxa. Requires Inoculation Fluid, REF R8325106. CE, 20 Panels/Kit The responsible firm name on the product labeling is Remel, 12076 Santa Fe Dr., Lenexa, KS 66215-3594
- Recall number
- Z-0870-2010
- Recalling firm
- Remel, Inc
- Product
- ERIC (Electronic RapID Compendium) Version 1.0.77, Catalog Number R8323600, used in conjunction with: remel RapID Yeast Plus System is a qualitative micromethod employing conventional and chromogenic substrates for the identification of medically important yeast, yeast-like, and related organisms isolated from human clinical specimens. A complete listing of the organisms addressed by the RapID Yeast Plus System is given in the RapID Yeast Plus Differential Chart. RapID" YEAST PLUS System, Catalog No. R8311007. 4 hour identification of yeast based on enzyme technology. Database includes over 40 taxa. Requires Inoculation Fluid, REF R8325106. CE, 20 Panels/Kit The responsible firm name on the product labeling is Remel, 12076 Santa Fe Dr., Lenexa, KS 66215-3594
- Status
- Terminated
- Initiated
- 2009-10-28
- Posted
- 2010-02-26
- Terminated
- 2010-03-03
- Reason
- There is an error in Version 1.0.77 of the electronic code compendium (ERIC) software. The error is limited to the database corresponding to the R8311007 RapID Yeast Plus System. The database erroneously includes Candida dubliniensis as well as displays an incorrect temperature range as an accessory test for codes with an overlap between Candida albicans and Candida dubliniensis. When using the
- Root cause
- Software design
- Action
- The firm issued an "IMPORTANT MEDICAL DEVICE PRODUCT RECALL NOTICE" dated October 29, 2009. After completion of the firm's investigation and risk assessment, Remel decided to notify customers of this situation and quarantine any remaining units on hand. The notification provided the consignees with instructions for obtaining and downloading the software patch. A response form was included with the recall letter and the consignee was asked to complete the form and return it to Remel. The consignees should contact the Technical Services Department at 800-255-6730 (USA) or 913-888-0939 (International), if they have inquiries concerning this advisory.
- Quantity
- 143 units of software
- Distribution
- Worldwide Distribution: USA, Europe and Canada
- Lot, serial or model codes
- ERIC (Electronic RapID Compendium) Version 1.0.77 -- Lot No. 0001096277
- 510(k) numbers
- ["K943621"]
- PMA numbers
- not on file
- Product code
- JXB
- Firm FEI
- 1924669
- Firm city
- Lenexa
- Firm state
- KS
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28