Becton Dickinson & Co.: BD BACTEC FX- Top Unit instrument, catalog number 441385, and Bottom Unit instrument, catalog 441386, available as a single or a stack configuration; Manufactured by BD Diagnostic Systems, 7 Loveton Circle, Sparks, MD 21152. The BD BACTEC FX instrument is designed for the rapid detection of bacteria and fungi in clinical specimens. Samples are drawn from patients and injected directly into BACTEC culture vials, which are placed into the instrument for incubation and testing.

Recall number
Z-0869-2014
Recalling firm
Becton Dickinson & Co.
Product
BD BACTEC FX- Top Unit instrument, catalog number 441385, and Bottom Unit instrument, catalog 441386, available as a single or a stack configuration; Manufactured by BD Diagnostic Systems, 7 Loveton Circle, Sparks, MD 21152. The BD BACTEC FX instrument is designed for the rapid detection of bacteria and fungi in clinical specimens. Samples are drawn from patients and injected directly into BACTEC culture vials, which are placed into the instrument for incubation and testing.
Status
Terminated
Initiated
2013-12-09
Posted
2014-01-30
Terminated
2014-08-20
Reason
Improperly functioning component of a diagnostic medical device may cause false negative results in clinical specimens.
Root cause
Nonconforming Material/Component
Action
BD Diagnostics initiated a field correction by phone on December 9, 2013, to affected customers and followed the calls with letters, e-mails, or fax. BD Diagnostics sent a secondary Urgent Field Corrective Action letter dated November 2013 to all affected customers. The recalling firm Secondary notifications were serial number specific and explained that some BD BACTEC FX instruments contain racks that may demonstrate variable indicator light intensities. Specifically, the Anonymous vial indicator lights which should appear yellow may instead appear more green in color. BD noted that affected instruments continue to function properly and will not report an incorrect result. The issue is limited to the yellow indicator lights only. Red and green indicator lights appear as intended. BD is developing software to correct this issue and anticipates that the new software will be available to begin instrument updates by February, 2014. The new software will not affect instrument if the indicator lights are functioning as intended. Until the software can be updated on your instrument, BD provided workflow instructions to ensure that the indicator light issue does not occur. For questions regarding this recall call 410-316-4054.
Quantity
198 units
Distribution
Worldwide Distribution - USA (nationwide) and Internationally to IN, AU, CN, AR, BE, PE, KR, BR, CO, SG, TH, BN, HK, JP, and CA.
Lot, serial or model codes
Catalog number 441385 and 441386
510(k) numbers
["K915796"]
PMA numbers
not on file
Product code
MDB
Firm FEI
1119779
Firm city
Sparks
Firm state
MD
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register