Skip to the content

FDA Medical Device Recalls (openFDA)

Deltec, Inc: Port-A-Cath II Low Profile Polysulfone/Titanium Venous Access System, with PolyFlow Polyurethane Catheter, 1.9mm O.D. x 1.0mm I.D., 6 Fr Introducer Set, REF 21-4083-24

Recall number
Z-0869-04
Recalling firm
Deltec, Inc
Product
Port-A-Cath II Low Profile Polysulfone/Titanium Venous Access System, with PolyFlow Polyurethane Catheter, 1.9mm O.D. x 1.0mm I.D., 6 Fr Introducer Set, REF 21-4083-24
Status
Terminated
Initiated
2004-03-29
Posted
2004-07-20
Terminated
2006-12-28
Reason
Some of the outer product pouches have voids in the seals and do not assure that the outer surfaces of the inner tray containing the port, catheter and procedural components remain sterile after sterilization.
Root cause
Other
Action
A recall letter dated March 26, 2004 gave customers the option of returning the products or using the products if the customer's procedures do not require sterility of the outer surface of the inner tray.
Quantity
not on file
Distribution
The recalled products were shipped nationwide in the United States and worldwide.
Lot, serial or model codes
Lots M26971 and M27045
510(k) numbers
["K962695"]
PMA numbers
not on file
Product code
LJT
Firm FEI
2183502
Firm city
Saint Paul
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register