Laerdal Medical (Suzhou) Co., Ltd.: Laerdal Compact Suction Unit (LCSU 4), Catalog Numbers 880051, 880061, 881151, 880052, 880062, 881152 - Product Usage: It is intended for intermittent operation to remove secretions, blood or vomit from a patient s airway to allow ventilation.

Recall number
Z-0868-2021
Recalling firm
Laerdal Medical (Suzhou) Co., Ltd.
Product
Laerdal Compact Suction Unit (LCSU 4), Catalog Numbers 880051, 880061, 881151, 880052, 880062, 881152 - Product Usage: It is intended for intermittent operation to remove secretions, blood or vomit from a patient s airway to allow ventilation.
Status
Terminated
Initiated
2020-12-18
Posted
not on file
Terminated
2023-12-11
Reason
Component failures may cause the unit to lose all suction. Users will need to obtain alternate suction devices or methods, leading to a delay in treatment and, in worst case scenario, potential suffocation if alternative suction or ventilation method is not administered by the user.
Root cause
Device Design
Action
On December 18, 2020, the firm sent Urgent Medical Device Recall letters to affected customers. Customers were informed that the service life of the LCSU 4 may be reduced. Customers were asked to identify and secure affected units. Customers were told that they could continue to use their LCSU 4 unit until they receive a replacement unit, but were provided with safety measures: - Follow local protocols for back-up suction devices. Do not repeatedly block the patient tube for more than 10 seconds at a time at vacuum levels greater than 300 mmHg. - Only use canisters, tubes and accessories that are approved for the device by Laerdal Medical. - When adjusting the vacuum level: When the desired vacuum level is reached; unblock the patient tube immediately. - During patient treatment do not occlude or block the patient tube by intentional bending or squeezing. The patient tube should only be temporarily occluded by suctioning matter. - When performing the Device Test: When the vacuum reaches minimum 500 mmHg immediately release the blocking. - If your affected LCSU 4 experiences a loss of peak suction, please report this to your local Laerdal Sales office, for US: 877-LAERDAL (523-7325), customerservice@laerdal.com or, for Canada: (416) 298-9600, savelives@laerdal.com Customers must complete the included reply form to acknowledge receipt of the recall notice and to identify any units needing replacement, and return the information to Stericycle.
Quantity
5035
Distribution
Worldwide distribution - US Nationwide distribution.
Lot, serial or model codes
Serial number range 1808090001 to 2006060045
510(k) numbers
["K120065"]
PMA numbers
not on file
Product code
JCX
Firm FEI
3004993040
Firm city
Suzhou
Firm state
not on file
Firm country
China

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register