FDA Medical Device Recalls (openFDA)
Vascular Solutions, Inc.: Vascular Solutions, SuperCross Microcatheter, Sterilized using ethylene oxide, Rx Only, Model numbers 5300 - 5307, 5324, 5325, 5329, 5332, 5336, 5340, and 5341. Product Usage: The SuperCross catheter is intended to be used in conjunction with steerable guidewires to access discrete regions of the coronary and/or peripheral vasculature. It may be used to facilitate placement and exchange of guidewires and other interventional devices and to subselectively infuse/deliver diagnostic and therapeutic agents
- Recall number
- Z-0868-2013
- Recalling firm
- Vascular Solutions, Inc.
- Product
- Vascular Solutions, SuperCross Microcatheter, Sterilized using ethylene oxide, Rx Only, Model numbers 5300 - 5307, 5324, 5325, 5329, 5332, 5336, 5340, and 5341. Product Usage: The SuperCross catheter is intended to be used in conjunction with steerable guidewires to access discrete regions of the coronary and/or peripheral vasculature. It may be used to facilitate placement and exchange of guidewires and other interventional devices and to subselectively infuse/deliver diagnostic and therapeutic agents
- Status
- Terminated
- Initiated
- 2013-01-29
- Posted
- 2013-02-20
- Terminated
- 2013-06-25
- Reason
- Vascular Solutions, Inc has discovered a potential problem with a number of lots of Supercross microcatheters. A potential for surface contamination on aged catheters was observed during in-house testing.
- Root cause
- Material/Component Contamination
- Action
- Vascular Solutions sent an "Urgent Medical device Recall" letter dated January 28, 29013 to all affected customers. The letter identified the affected product, described the problem and actions to be taken. The letter advised customers to remove the product from their inventory and to complete and return the VSI Account Inventory Form. For questions contact your local Sales Representative.
- Quantity
- 4731
- Distribution
- Worldwide Distribution - USA Nationwide including the states of AL, AZ, AR, CA, CO, CT, DE, CD, FL, GA, HI, ID, IL ,IN, IA, KY, LA, ME, MD, MA, MI,MN, MO, MT, NE, NV, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, SC, SD, TN, TX, UT, VT VA, WA, WV, WI, WY and AUSTRIA, CANDA, DENMARK, FINLAND, FRANCE, GERMANY, GREECE, HUNGARY, ISRAEL,. ITALY, MEXICO, NETHERLANDS, NORWAY, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, TURKEY, UNITED KINGDOM.
- Lot, serial or model codes
- 559450, 551750, 552370, 552761, 552840, 552954, 553267, 553647, 553931, 554269, 554270, 554413. 555958, 557192, 557824, 558222, 558894, 559827, 560215, 551751, 552371, 552841, 553090, 553218, 553434, 553529, 554029, 554268, 554271, 555759, 555953, 557251, 559570, 559981, 560253, 560079, 560080, 554123, 556182, 558905, 560114, 554210, 555991, 556842, 557253, 554209, 555223, 555992, 556536, 556747, 557124, 558327, 558991, 560115, 561206, 554208, 555490, 555993, 556674, 556843, 558018, 558798, 559707, 560433, 561207, 554040, 555007, 555224, 555358, 558542, 560434, 561459, 554122, 555008, 555225, 555994, 558992, 561208, 560221, 560222, 560223, 560995, 560996.
- 510(k) numbers
- ["K101659"]
- PMA numbers
- not on file
- Product code
- DQY
- Firm FEI
- 3002827704
- Firm city
- Maple Grove
- Firm state
- MN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28