Stryker Endoscopy: Dri-Lok Disposable Cannula, Part numbers 3910-075-500, 3910-075-501, 3910-075-502, 3910-075-650, 3910-075-651, 3910-075-652, 3910-075-800, 3910-075-801, 3910-075-802, 3910-090-500, 3910-090-502, 3910-090-650, 3910-090-652, 3910-090-800, and 3910-090-802. The obturator is used to create an opening into the patient's joint and is removed after the cannula is inserted. The cannula then allows instruments and scopes access to the joint while maintaining a seal against the loss of saline fluid used to keep the joint clean. The seal is formed by two elastomeric valves. The first controls loss of water when there is no instrument or scope therein, while the second provides a seal when the cannula is in use. The valves operate automatically. Manufactured by Stryker Endoscopy Puerto Rico
- Recall number
- Z-0868-2009
- Recalling firm
- Stryker Endoscopy
- Product
- Dri-Lok Disposable Cannula, Part numbers 3910-075-500, 3910-075-501, 3910-075-502, 3910-075-650, 3910-075-651, 3910-075-652, 3910-075-800, 3910-075-801, 3910-075-802, 3910-090-500, 3910-090-502, 3910-090-650, 3910-090-652, 3910-090-800, and 3910-090-802. The obturator is used to create an opening into the patient's joint and is removed after the cannula is inserted. The cannula then allows instruments and scopes access to the joint while maintaining a seal against the loss of saline fluid used to keep the joint clean. The seal is formed by two elastomeric valves. The first controls loss of water when there is no instrument or scope therein, while the second provides a seal when the cannula is in use. The valves operate automatically. Manufactured by Stryker Endoscopy Puerto Rico
- Status
- Terminated
- Initiated
- 2008-12-01
- Posted
- 2009-01-16
- Terminated
- 2011-01-10
- Reason
- Leak: Leak-resistant barrier may not properly seal and leak when an instrument is inserted, and keep leaking after instrument is removed.
- Root cause
- Component change control
- Action
- Recall initiated 12/01/2008, Sales representatives were notified via telephone and e-mail. Product advisories entitled ""Urgent: Device Removal" and dated December 1, 2008, were sent to all consignees. All unused product is to be returned to the firm. For further information, please contact Stryker Endoscopy by telephone at 408-754-2124.
- Quantity
- 4,194 boxes (24,750 units)
- Distribution
- Worldwide Distribution --- including USA and countries of Australia, Brazil, Canada, Chile, China, France, Germany, Greece, India, Italy, New Zealand, Poland, Portugal, Sweden, Spain, Singapore, Switzerland, and UK.
- Lot, serial or model codes
- Lot numbers 08135AG2 through 08294AG2.
- 510(k) numbers
- ["K944210"]
- PMA numbers
- not on file
- Product code
- HRX
- Firm FEI
- 2936485
- Firm city
- San Jose
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28