Stryker Endoscopy: Dri-Lok Disposable Cannula, Part numbers 3910-075-500, 3910-075-501, 3910-075-502, 3910-075-650, 3910-075-651, 3910-075-652, 3910-075-800, 3910-075-801, 3910-075-802, 3910-090-500, 3910-090-502, 3910-090-650, 3910-090-652, 3910-090-800, and 3910-090-802. The obturator is used to create an opening into the patient's joint and is removed after the cannula is inserted. The cannula then allows instruments and scopes access to the joint while maintaining a seal against the loss of saline fluid used to keep the joint clean. The seal is formed by two elastomeric valves. The first controls loss of water when there is no instrument or scope therein, while the second provides a seal when the cannula is in use. The valves operate automatically. Manufactured by Stryker Endoscopy Puerto Rico

Recall number
Z-0868-2009
Recalling firm
Stryker Endoscopy
Product
Dri-Lok Disposable Cannula, Part numbers 3910-075-500, 3910-075-501, 3910-075-502, 3910-075-650, 3910-075-651, 3910-075-652, 3910-075-800, 3910-075-801, 3910-075-802, 3910-090-500, 3910-090-502, 3910-090-650, 3910-090-652, 3910-090-800, and 3910-090-802. The obturator is used to create an opening into the patient's joint and is removed after the cannula is inserted. The cannula then allows instruments and scopes access to the joint while maintaining a seal against the loss of saline fluid used to keep the joint clean. The seal is formed by two elastomeric valves. The first controls loss of water when there is no instrument or scope therein, while the second provides a seal when the cannula is in use. The valves operate automatically. Manufactured by Stryker Endoscopy Puerto Rico
Status
Terminated
Initiated
2008-12-01
Posted
2009-01-16
Terminated
2011-01-10
Reason
Leak: Leak-resistant barrier may not properly seal and leak when an instrument is inserted, and keep leaking after instrument is removed.
Root cause
Component change control
Action
Recall initiated 12/01/2008, Sales representatives were notified via telephone and e-mail. Product advisories entitled ""Urgent: Device Removal" and dated December 1, 2008, were sent to all consignees. All unused product is to be returned to the firm. For further information, please contact Stryker Endoscopy by telephone at 408-754-2124.
Quantity
4,194 boxes (24,750 units)
Distribution
Worldwide Distribution --- including USA and countries of Australia, Brazil, Canada, Chile, China, France, Germany, Greece, India, Italy, New Zealand, Poland, Portugal, Sweden, Spain, Singapore, Switzerland, and UK.
Lot, serial or model codes
Lot numbers 08135AG2 through 08294AG2.
510(k) numbers
["K944210"]
PMA numbers
not on file
Product code
HRX
Firm FEI
2936485
Firm city
San Jose
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register