Gsi Group Inc: PRN 50-M+, GE part number 2062759-001, manufacturer part number 600-23310-01 Product Usage: The PRN 50M+ Paper Chart Recorder thermally records patient data on a paper strip. This device is intended for use under direct supervision of a licensed health care practitioner. This device is not intended for home use.

Recall number
Z-0866-2016
Recalling firm
Gsi Group Inc
Product
PRN 50-M+, GE part number 2062759-001, manufacturer part number 600-23310-01 Product Usage: The PRN 50M+ Paper Chart Recorder thermally records patient data on a paper strip. This device is intended for use under direct supervision of a licensed health care practitioner. This device is not intended for home use.
Status
Terminated
Initiated
2015-12-02
Posted
not on file
Terminated
2018-06-06
Reason
The item was recalled due to an increased user/patient risk of exposure to fire and/or electrical shock due to the failure of the PRN 50-M+ Digital Writer
Root cause
Nonconforming Material/Component
Action
On December 2, 2015 GSI Group Corp. distributed Medical Device Correction Notice letter and confirmation forms dated November 18, 2015 to their customers via FedEx. GSI Group Corp. warned customers about a potential safety issue associated with the flammability and fluid ingress rating of the outer enclosure material. Customers are advised to follow all intended usage, cleaning and maintenance instructions that are found in the PRN 50-M+ Digital Writer Technical Manual. It is recommended that customers complete and return the attached Medical Device Correction Confirmation form via email to customerservice@insidesales-solutions.com. GSI Group Corp. will provide a correction at no charge. Once the confirmation form is received, GSI Group Corp. will make arragements to send replacements. Customers with questions can call 781-266-5700.
Quantity
2257 units
Distribution
Worldwide Distribution - US Nationwide including Puerto Rico; Internationally: Australia, Bahamas, Canada, Germany, Finland, France, Great Britain, Hong Kong, India, Ireland, Italy, Japan, South Korea, Mauritus, Mexico, Peru, Saudi Arabia, Singapore, Taiwan, United Arab Emirates & United Kingdom
Lot, serial or model codes
Serial Number range: 1801405 through 1918761
510(k) numbers
not on file
PMA numbers
not on file
Product code
DSF
Firm FEI
3007083974
Firm city
Bedford
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register