Merit Medical Systems, Inc.: Merit Medical Systems Inc., Resolve Biliary Locking Drainage Catheter, Models RBC and RBDC, used for drainage of bile within the biliary system.

Recall number
Z-0866-2013
Recalling firm
Merit Medical Systems, Inc.
Product
Merit Medical Systems Inc., Resolve Biliary Locking Drainage Catheter, Models RBC and RBDC, used for drainage of bile within the biliary system.
Status
Terminated
Initiated
2013-01-16
Posted
2013-02-20
Terminated
2015-09-16
Reason
A consumer complaint that a catheter was observed to be in several pieces under patient fluoroscopic visualization.
Root cause
Nonconforming Material/Component
Action
Merit initiated the recall on January 16, 2013. A voicemail script notification to users and Q&A document was provided to those sales representative that had received and or distributed product. On January 16, 2013, Merit sales representatives began contacting each physician and/or user facility that had been provided product. Written communication distribution began on 22 January 2013. Affected sales representatives, physicians, and hospitals were provided written identification information, instruction to not distribute or continue placing these devices, and on how to return any unused product if applicable. All product on hand at Merit was inventoried and remains under quarantined status. Further questions please call (801) 208-4301,
Quantity
387 units
Distribution
Worldwide Distribution - USA including the states of NJ, FL, GA, MO, AZ, UT and the country of Denmark
Lot, serial or model codes
Lot numbers H399931S1; H411376; H399938S1; H411373; H399940S1; H411374; H399942S1; H411375; H399948S1; H411380; H399950S1; H411377; H39952S1; H411378; H39953S1; H411379.
510(k) numbers
["K121832"]
PMA numbers
not on file
Product code
FGE
Firm FEI
1721504
Firm city
South Jordan
Firm state
UT
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register