Skip to the content

FDA Medical Device Recalls (openFDA)

Halyard Health: KimVent Microcuff Subglottic Suctioning, Endotracheal Tube, 7.0 mm, Distributed by Kimberly Clark Global Sales, LLC. Used for airway management by oral intubation of the trachea and for removal of secretions that accumulate in the subglottic space.

Recall number
Z-0865-2015
Recalling firm
Halyard Health
Product
KimVent Microcuff Subglottic Suctioning, Endotracheal Tube, 7.0 mm, Distributed by Kimberly Clark Global Sales, LLC. Used for airway management by oral intubation of the trachea and for removal of secretions that accumulate in the subglottic space.
Status
Terminated
Initiated
2014-11-17
Posted
2014-12-24
Terminated
2015-08-18
Reason
The cuff inflation line may detach from the endotracheal tube during use.
Root cause
Nonconforming Material/Component
Action
The firm, Halyard Health, sent an "URGENT: VOLUNTARY PRODUCT RECALL" letter dated November 17, 2014 via FedEx to its Consignees/customers. The letter describes the product, problem and actions to be taken. 'the customers were instructed to please evaluate your facilities for unused inventory of KimVent* Microcuff* Subglottic Suctioning Endotracheal Tubes, if present, quarantine and discontinue use; complete the attached Distributor Product Recall Response Form via fax to: 678-254-0391 within five (5) business days of receipt of this letter; if you are in possession of the affected product lots, a Halyard Health representative will provide a return authorization and instructions on shipping product to a specified distribution center, and; follow your internal product recall procedures to identify and address affected product. If you require further assistance, please call Halyard Health Regulatory Affairs at 770-587-7400.
Quantity
250 cases (10 tubes per case)
Distribution
Worldwide Distribution: US (nationwide) AL, AR, AZ, CA, CT, FL, GA, IN, KS, KY, MA, ME, MI, MO, MS, MT, ND, NH, OH, OK, OR, PA, TN, UT, VA, and WV; and countries of: Australia, Canada, Europe (i.e., France, Germany, Great Britain, Ireland, Italy, Netherlands, Portugal, Sweden, & Switzerland), Hong Kong, New Zealand, and United Arab Emirates.
Lot, serial or model codes
Product Code: 13220, Lot numbers: AA3213, AA3227, AA3227V01, AA3245, AA3245V02, AA3253V02, AA3269V01, AA3274V01, AA4038, AA4038V03, AA4038V04, AA4038V06, AA4062, AA4062V01, AA4104V02, AA4111V01, AA4113V01, AA4134V01 & AA4146V01, AA4146V04, AA4174V01.
510(k) numbers
["K131254"]
PMA numbers
not on file
Product code
BTR
Firm FEI
3011157493
Firm city
Roswell
Firm state
GA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register