Skip to the content

FDA Medical Device Recalls (openFDA)

Siemens Medical Solutions USA, Inc.: Acuson Diagnostic Ultrasound System Transducer; Model Number(s): L582 Linear Array Transducer for use on the 128XP Ultrasound System; Catalog Numbers: 23656 (New), 45441 (Refurb)

Recall number
Z-0865-03
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
Acuson Diagnostic Ultrasound System Transducer; Model Number(s): L582 Linear Array Transducer for use on the 128XP Ultrasound System; Catalog Numbers: 23656 (New), 45441 (Refurb)
Status
Terminated
Initiated
2003-03-31
Posted
2003-05-22
Terminated
2003-09-22
Reason
The product has a software condition in that the surface temperature of the transducer may reach above the thermal limit.
Root cause
Other
Action
On 3/24/03, the firm issued letters via US Mail to all its direct consignees informing them of the affected device and providing instructions on the recall.
Quantity
220 units
Distribution
Nationwide and worldwide. The product received nationwide distribution to 123 consignees: 23 Distributors (50 transducers) and consumers/users: approximately 100 customers (146 transducers). 2 transducers were consigned to Federal Government locations: (1) Martin Army Community Hospital, 9200 Marne Rd., Fort Benning, GA 31905, (2) Lackland AFB, Wilford Hall, 2200 Bergquist Dr., #100, Lackland AFB, TX 78236. The firm has also distributed product into international channels through distributors. Countries include, Austria; Barbados; Brazil; Canada; China (PRC); Egypt, France, Germany; Greece; India; Italy; Japan; South Korea; Netherlands; Russia; Saudi Arabia; Sweden; Taiwan; and United Kingdom. There is no Canadian distribution. The firm is continuing to obtain the international distribution through the distributors. The recall was appropriately extended to the consumer/user level; i.e., the wholesalers/distributors, and users who received the recalled product.
Lot, serial or model codes
All Serial Numbers: XXXXXXXX (8-digit)
510(k) numbers
["K884632"]
PMA numbers
not on file
Product code
MAA
Firm FEI
2936884
Firm city
Mountain View
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register