FDA Medical Device Recalls (openFDA)
Siemens Medical Solutions USA, Inc.: Acuson Diagnostic Ultrasound System Transducer; Model Number(s): L582 Linear Array Transducer for use on the 128XP Ultrasound System; Catalog Numbers: 23656 (New), 45441 (Refurb)
- Recall number
- Z-0865-03
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- Acuson Diagnostic Ultrasound System Transducer; Model Number(s): L582 Linear Array Transducer for use on the 128XP Ultrasound System; Catalog Numbers: 23656 (New), 45441 (Refurb)
- Status
- Terminated
- Initiated
- 2003-03-31
- Posted
- 2003-05-22
- Terminated
- 2003-09-22
- Reason
- The product has a software condition in that the surface temperature of the transducer may reach above the thermal limit.
- Root cause
- Other
- Action
- On 3/24/03, the firm issued letters via US Mail to all its direct consignees informing them of the affected device and providing instructions on the recall.
- Quantity
- 220 units
- Distribution
- Nationwide and worldwide. The product received nationwide distribution to 123 consignees: 23 Distributors (50 transducers) and consumers/users: approximately 100 customers (146 transducers). 2 transducers were consigned to Federal Government locations: (1) Martin Army Community Hospital, 9200 Marne Rd., Fort Benning, GA 31905, (2) Lackland AFB, Wilford Hall, 2200 Bergquist Dr., #100, Lackland AFB, TX 78236. The firm has also distributed product into international channels through distributors. Countries include, Austria; Barbados; Brazil; Canada; China (PRC); Egypt, France, Germany; Greece; India; Italy; Japan; South Korea; Netherlands; Russia; Saudi Arabia; Sweden; Taiwan; and United Kingdom. There is no Canadian distribution. The firm is continuing to obtain the international distribution through the distributors. The recall was appropriately extended to the consumer/user level; i.e., the wholesalers/distributors, and users who received the recalled product.
- Lot, serial or model codes
- All Serial Numbers: XXXXXXXX (8-digit)
- 510(k) numbers
- ["K884632"]
- PMA numbers
- not on file
- Product code
- MAA
- Firm FEI
- 2936884
- Firm city
- Mountain View
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28