Skip to the content

FDA Medical Device Recalls (openFDA)

Philips Ultrasound Inc: All EPIQ and Affiniti Ultrasound systems. Model numbers - EPIQ 5G, EPIC 5C, EPIQ 5W, EPIQ 7G, EPIC 7C, EPIQ 7W, EPIQ CVx, Affiniti 30, Affiniti 50, Affiniti 70. Diagnostic Ultrasound System. MOD - EPIQ 5 GMDN - 40761 100-240V 50/60 Hz, 600 VA MOD - EPIQ 7 GMDN - 40761 100-240V 50/60 Hz, 600 VA MOD - EPIQ 100-240V 50/60 Hz, 600 VA MOD - Affinity 30 GMDN - 40761 100-240V 50/60 Hz, 450 VA MOD - Affinity 50 GMDN - 40761 100-240V 50/60 Hz, 450 VA MOD - Affinity 70 GMDN - 40761 100-240V 50/60 Hz, 450 VA

Recall number
Z-0864-2021
Recalling firm
Philips Ultrasound Inc
Product
All EPIQ and Affiniti Ultrasound systems. Model numbers - EPIQ 5G, EPIC 5C, EPIQ 5W, EPIQ 7G, EPIC 7C, EPIQ 7W, EPIQ CVx, Affiniti 30, Affiniti 50, Affiniti 70. Diagnostic Ultrasound System. MOD - EPIQ 5 GMDN - 40761 100-240V 50/60 Hz, 600 VA MOD - EPIQ 7 GMDN - 40761 100-240V 50/60 Hz, 600 VA MOD - EPIQ 100-240V 50/60 Hz, 600 VA MOD - Affinity 30 GMDN - 40761 100-240V 50/60 Hz, 450 VA MOD - Affinity 50 GMDN - 40761 100-240V 50/60 Hz, 450 VA MOD - Affinity 70 GMDN - 40761 100-240V 50/60 Hz, 450 VA
Status
Terminated
Initiated
2020-10-27
Posted
not on file
Terminated
2023-04-19
Reason
The manufacturer has determined that with certain uncommon workflows there is potential for incorrect patient data to be displayed and saved into an exam. The issue(s) manifest differently for different versions of software.
Root cause
Software design
Action
On 10/27/2020,the manufacturer/ recalling firm started sending an Urgent Medical Device Correction letter (MDC) with recommended workflows to prevent the potential issues. Philips is asking customers to follow the Action to be Taken by Customer/User section of the Urgent Medical Device Correction letter. The effectiveness check will include the delivery confirmation of the Urgent Medical Device Correction Letter in the USA or the Field Safety Notice (FSN) worldwide. In addition, effectiveness checks will be conducted by confirmation that the customer has received the letter and have read/understood the actions documented within (customer reply form).
Quantity
57,430 units
Distribution
Worldwide distribution - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV and WY. The countries of Afghanistan, Albania, Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bermuda, Bolivia, Bosnia and Herzegovina, Brazil, Brunei Darussalam, Bulgaria, Cambodia, Canada, Chile, China, Colombia, Cook Islands, Costa Rica, Corte D'Ivoire, Croatia, Cuba, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Estonia, Ethiopia, Faroe Islands, Finland, France, French Guiana, French Polynesia, Georgia, Germany, Ghana, Gibraltar, Greece, Greenland, Guadeloupe, Guatemala, Guernsey, Honduras, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Korea, Republic of, Kuwait, Kyrgyzstan, Lao People's Democratic Republic, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Macao, Macedonia, Malaysia, Maldives, Mali, Malta, Martinique, Mauritius, Mayotte, Mexico, Moldova, Monaco, Mongolia, Montenegro, Morocco, Myanmar, Namibia, Nepal, Netherlands, New Caledonia, New Zealand, Nicaragua, Nigeria, Norway, Oman, Pakistan, Palestine, State of, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Puerto Rico, Qatar, R¿union, Romania, Russian Federation, Saudi Arabia, Senegal, Serbia, Sierra Leone, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Suriname, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Tanzania, Thailand, Tunisia, Turkey, Turkmenistan, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Uzbekistan, Venezuela, Viet Nam, Virgin Islands (British), Virgin Islands (U.S.), Yemen, Zimbabwe.
Lot, serial or model codes
All EPIQ and Affiniti Ultrasound systems with software versions 1.0, 1.1.x, 1.2.x, 1.3.x, 1.4.x, 1.5.x, 1.8.x, 2.0.x, 3.0.x, 4.0.x, 5.0, 5.0.1
510(k) numbers
["K182857"]
PMA numbers
not on file
Product code
IYN
Firm FEI
3019216
Firm city
Bothell
Firm state
WA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28