Codman & Shurtleff, Inc.: Codman Certas Programmable Valve In Line Valve with Siphonguard, Unitized Catheter and Accessories Product Code: 82-8806 Product Usage: The CODMAN CERTAS Programmable Valve is an implantable device that provides constant intraventricular pressure and drainage of CSF (cerebrospinal fluid) for the management of hydrocephalus. The valve can be set to a choice of eight opening pressure settings for constant intraventricular perssure drainage of CSF. The CODMAN CERTAS Therapy Management System (TMS) allows the non-invasive reading or adjustment of the valve setting.

Recall number
Z-0863-2014
Recalling firm
Codman & Shurtleff, Inc.
Product
Codman Certas Programmable Valve In Line Valve with Siphonguard, Unitized Catheter and Accessories Product Code: 82-8806 Product Usage: The CODMAN CERTAS Programmable Valve is an implantable device that provides constant intraventricular pressure and drainage of CSF (cerebrospinal fluid) for the management of hydrocephalus. The valve can be set to a choice of eight opening pressure settings for constant intraventricular perssure drainage of CSF. The CODMAN CERTAS Therapy Management System (TMS) allows the non-invasive reading or adjustment of the valve setting.
Status
Terminated
Initiated
2013-05-16
Posted
2014-01-29
Terminated
2015-06-02
Reason
Codman Certas Programmable Valves used for hydrocephalus may not operate properly
Root cause
Device Design
Action
Codman issued an Urgent Notice dated May 16, 2013, to all affected customers. The letter identifed the affected product, problem and actions to be taken. Mitigation instructions are provided and if not followed or are not successful, either issue could lead to symptoms associated with under- or over-drainage. Units not implanted are to be returned. A U.S Acknowledgement Form and fax are requested to be completed and faxed to: 1-888-943-4193, or e-mail the form to codman8894@stericycle.com. If you have any questions on concerns in regards to this recall, please contact your Codman Representative or Stericycle at 1-888-943-2395.
Quantity
409 units
Distribution
Worldwide Distribution - US Nationwide, Australia, Austria, Belgium, Brazil, Switzerland, Czech Republic, Germany, Spain, Finland, France, Great Britain, Israel, Italy, Netherlands, Norway, Portugal, Russia, Saudi Arabia, Sweden, Slovenia, Slovakia, Turkey, South Africa, and UK.
Lot, serial or model codes
Lot Codes:  CMJCMD CMBC2Z CMDB7C CMJC49 CMBCPR CMCC2T CMDBHC CMLBLZ CMMB14 CNDBCY CNDCV2 CNJCFT CNKBH6 CNLB57 CNMDNL CNNCWD CNPB1Z CPBB3J CPBDBB CPCCM0 CPDBYK
510(k) numbers
["K112156","K113526"]
PMA numbers
not on file
Product code
JXG
Firm FEI
3002947440
Firm city
Raynham
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register