Beckman Coulter Inc.: Beckman Coulter UniCel DxC Synchron Clinical System ISE Flow Cell. PN Numbers: A10405 (UniCel DxC 600); A11810 (UniCel DxC 600 PRO); A27318 (UniCel DxC 600i); A11816 (UniCel DxC 800); A11812 (UniCel DxC 800 PRO); A59102 (UniCel DxC 880i); A64903 (UniCel DxC 680i); A64871 (UniCel DxC 660i); A64871 (UniCel DxC 860i). Intended for the in vitro determination of a variety of general chemistries, therapeutic drugs, and other chemistries of clinical interest in biological fluids such as serum, plasma, urine, or cerebrospinal fluid.

Recall number
Z-0863-2010
Recalling firm
Beckman Coulter Inc.
Product
Beckman Coulter UniCel DxC Synchron Clinical System ISE Flow Cell. PN Numbers: A10405 (UniCel DxC 600); A11810 (UniCel DxC 600 PRO); A27318 (UniCel DxC 600i); A11816 (UniCel DxC 800); A11812 (UniCel DxC 800 PRO); A59102 (UniCel DxC 880i); A64903 (UniCel DxC 680i); A64871 (UniCel DxC 660i); A64871 (UniCel DxC 860i). Intended for the in vitro determination of a variety of general chemistries, therapeutic drugs, and other chemistries of clinical interest in biological fluids such as serum, plasma, urine, or cerebrospinal fluid.
Status
Terminated
Initiated
2010-01-04
Posted
2010-03-04
Terminated
2012-04-17
Reason
Excessive Buildup of protein, bacteria and sample tube additives in the ISE flow cell may cause erroneous NA (sodium) results.
Root cause
Device Design
Action
An Urgent: Product Corrective Action letter dated January 4, 2010, was mailed to customers advising of erroneous sodium results derived from build up of protein, bacteria and additives in the ISE flow cell. The letter discussed the impact and actions to be taken along with some additional recommendations. Customers are advised to follow the new maintenance instructions included with the letter. The information in the letter should be shared with laboratory staff and the notification should be retained as part of the Quality System documentation. A copy of the letter should be provided to any other laboratory that affected product was forwarded to. The enclosed response form should be completed and returned. Questions should be directed to the Customer Support Center at 1-800-854-3633.
Quantity
2630
Distribution
Worldwide Distribution -- USA and Canada.
Lot, serial or model codes
All serial numbers. PN Numbers: A10405 (UniCel DxC 600); A11810 (UniCel DxC 600 PRO); A27318 (UniCel DxC 600i); A11816 (UniCel DxC 800); A11812 (UniCel DxC 800 PRO); A59102 (UniCel DxC 880i); A64903 (UniCel DxC 680i); A64871 (UniCel DxC 660i); A64871 (UniCel DxC 860i); and A64935 (UniCel DxC 860i).
510(k) numbers
["K042291","K060256"]
PMA numbers
not on file
Product code
JJE
Firm FEI
2050012
Firm city
Brea
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register