Skip to the content

FDA Medical Device Recalls (openFDA)

Miltex, Inc.: Stainless Steel Surgical Blades Part no 4-312B. The product is supplied sterile individually packaged in boxes containing 100 blades.

Recall number
Z-0862-2009
Recalling firm
Miltex, Inc.
Product
Stainless Steel Surgical Blades Part no 4-312B. The product is supplied sterile individually packaged in boxes containing 100 blades.
Status
Terminated
Initiated
2008-11-17
Posted
2009-02-20
Terminated
2009-02-23
Reason
Blade may break: Over milling of the blade resulted in tip more narrow than specified.
Root cause
Process control
Action
An Urgent Medical Device Recall letter dated 11/17/08 with an accompanying fax back form was sent to each customer. Each distributor was instructed to review their inventory, complete the form, and return any affected product. The distributors were also instructed to conduct a sub recall notification.
Quantity
200 boxes
Distribution
Nationwide Distribution and Canada.
Lot, serial or model codes
Lot number S07C06
510(k) numbers
not on file
PMA numbers
not on file
Product code
GES
Firm FEI
2518040
Firm city
York
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register