Skip to the content

FDA Medical Device Recalls (openFDA)

Konica Minolta Medical Imaging USA, Inc.: REGIUS Image Manager (AKA REGIUS IM, RIM). Catalogue Number 5900554. Centralized CR Image Quality Control. Patient throughput optimization via offloaded processing from various CR readers. Technologist review and audit. Key processing for specialized exams, totally automated processing for hands off operation. Centralized workflow device for dispersed readers environments. HIS/RIS patient status updates. Demographic Correction.

Recall number
Z-0862-06
Recalling firm
Konica Minolta Medical Imaging USA, Inc.
Product
REGIUS Image Manager (AKA REGIUS IM, RIM). Catalogue Number 5900554. Centralized CR Image Quality Control. Patient throughput optimization via offloaded processing from various CR readers. Technologist review and audit. Key processing for specialized exams, totally automated processing for hands off operation. Centralized workflow device for dispersed readers environments. HIS/RIS patient status updates. Demographic Correction.
Status
Terminated
Initiated
2006-02-06
Posted
2006-05-16
Terminated
2008-08-22
Reason
If the patient demographic data is entered into the CS-1 manually, and the user fails to enter any data in one of the demographic data fields labeled as Birthday, Sex, Name or Patient Comments, etc, the RIM will populate the fields where no data has been entered with data from the previous patient.
Root cause
Other
Action
The firm phoned the customers on 2/6/2006 to inform them of the recall and the corrective action associated to the recall. This will be followed up with a fax, e-mail and letter. The final step will be visiting the location to install the correct software.
Quantity
58 units
Distribution
The units were distributed to hospitals nationwide through the firm''s sales force and resellers. There are no government accounts.
Lot, serial or model codes
58 units distributed between June 2003 and January 2006. Software version prior to January 31, 2006.
510(k) numbers
["K051523"]
PMA numbers
not on file
Product code
LLZ
Firm FEI
2241281
Firm city
Wayne
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register