PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.: Azurion 7 M20 FlexArm system (Software version R2.x) with Maquet Magnus Table (Software version 05.02.12)
- Recall number
- Z-0861-2024
- Recalling firm
- PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
- Product
- Azurion 7 M20 FlexArm system (Software version R2.x) with Maquet Magnus Table (Software version 05.02.12)
- Status
- Open, Classified
- Initiated
- 2023-12-08
- Posted
- 2024-01-30
- Terminated
- not on file
- Reason
- Loss of connectivity between the FlexArm and the Table due to a software issue related to the date/time format. This connection issue leads to a loss of movement of the arm or table.
- Root cause
- Under Investigation by firm
- Action
- The recalling firm issued an URGENT: MEDICAL DEVICE CORRECTION notification dated 12/08/2023 by certified mail to consignees. Consignees with affected devices are asked to utilize the provided workaround instructions. Additionally, consignees are asked to make all users aware of this recall notification and keep a copy of this notification with system documentation. Getinge Field Service Engineers will be dispatched to the field to install software updates to affected devices. Consignees with any questions can contact the recalling firm by phone at 1-800-722-9377. Consignees are asked to return the provided response form to the recalling firm at ITG_Recalls@philips.com.
- Quantity
- 108 systems (14 US, 1 Canada, 93 ROW)
- Distribution
- Worldwide - US Nationwide distribution in the states of CO, FL, IL, MS, NV and TX. The country of Canada.
- Lot, serial or model codes
- (1) System Code: 722079; (1) UDI/DI: 00884838085268; (1) Serial Number: 1005, 866, 1361, 952, 1241, 1169, 756, 703351, 1135, 1122, 824, 748, 630, 638, 703385, 812, 703346, 1303, 1313, 893, 768, 929, 703638, 1357, 703810, 934, 1202, 703461, 703706, 1172, 703959, 703845, 703705; (2) System Code: 722224; (2) UDI/DI: 00884838099258; (2) Serial Number: 539, 395, 278, 134, 730, 252, 370, 36, 572, 645, 272, 495, 781, 62, 1388, 352, 1655, 690, 555, 895, 2165, 861, 60, 20, 648, 2231, 1778, 1431, 778, 477, 917, 1718, 388, 2042, 150, 764, 2036, 1421, 157, 879, 212, 144, 1463, 347, 575, 2187, 355, 2576, 1332, 424, 387, 863, 211, 474, 712, 549, 2309, 799, 2283, 639, 1700, 2071, 537, 1589, 2208, 2057, 771, 772, 1216, 1667;
- 510(k) numbers
- ["K200917"]
- PMA numbers
- not on file
- Product code
- OWB
- Firm FEI
- 3002807880
- Firm city
- Best
- Firm state
- not on file
- Firm country
- Netherlands
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28