Skip to the content

FDA Medical Device Recalls (openFDA)

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.: Azurion 7 M20 FlexArm system (Software version R2.x) with Maquet Magnus Table (Software version 05.02.12)

Recall number
Z-0861-2024
Recalling firm
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Product
Azurion 7 M20 FlexArm system (Software version R2.x) with Maquet Magnus Table (Software version 05.02.12)
Status
Open, Classified
Initiated
2023-12-08
Posted
2024-01-30
Terminated
not on file
Reason
Loss of connectivity between the FlexArm and the Table due to a software issue related to the date/time format. This connection issue leads to a loss of movement of the arm or table.
Root cause
Under Investigation by firm
Action
The recalling firm issued an URGENT: MEDICAL DEVICE CORRECTION notification dated 12/08/2023 by certified mail to consignees. Consignees with affected devices are asked to utilize the provided workaround instructions. Additionally, consignees are asked to make all users aware of this recall notification and keep a copy of this notification with system documentation. Getinge Field Service Engineers will be dispatched to the field to install software updates to affected devices. Consignees with any questions can contact the recalling firm by phone at 1-800-722-9377. Consignees are asked to return the provided response form to the recalling firm at ITG_Recalls@philips.com.
Quantity
108 systems (14 US, 1 Canada, 93 ROW)
Distribution
Worldwide - US Nationwide distribution in the states of CO, FL, IL, MS, NV and TX. The country of Canada.
Lot, serial or model codes
(1) System Code: 722079; (1) UDI/DI: 00884838085268; (1) Serial Number: 1005, 866, 1361, 952, 1241, 1169, 756, 703351, 1135, 1122, 824, 748, 630, 638, 703385, 812, 703346, 1303, 1313, 893, 768, 929, 703638, 1357, 703810, 934, 1202, 703461, 703706, 1172, 703959, 703845, 703705; (2) System Code: 722224; (2) UDI/DI: 00884838099258; (2) Serial Number: 539, 395, 278, 134, 730, 252, 370, 36, 572, 645, 272, 495, 781, 62, 1388, 352, 1655, 690, 555, 895, 2165, 861, 60, 20, 648, 2231, 1778, 1431, 778, 477, 917, 1718, 388, 2042, 150, 764, 2036, 1421, 157, 879, 212, 144, 1463, 347, 575, 2187, 355, 2576, 1332, 424, 387, 863, 211, 474, 712, 549, 2309, 799, 2283, 639, 1700, 2071, 537, 1589, 2208, 2057, 771, 772, 1216, 1667;
510(k) numbers
["K200917"]
PMA numbers
not on file
Product code
OWB
Firm FEI
3002807880
Firm city
Best
Firm state
not on file
Firm country
Netherlands

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register