Siemens Medical Solutions USA, Inc: Siemens AXIOM Aristos FX Plus solid state x-ray imager. Image-intensified fluoroscopic x-ray system.

Recall number
Z-0860-2012
Recalling firm
Siemens Medical Solutions USA, Inc
Product
Siemens AXIOM Aristos FX Plus solid state x-ray imager. Image-intensified fluoroscopic x-ray system.
Status
Terminated
Initiated
2011-12-12
Posted
2012-01-25
Terminated
2014-01-14
Reason
Fixation bolts for the mounting of the X-ray tube holder became mixed with an incorrect bolt type.
Root cause
Process control
Action
Siemens Healthcare sent a "SAFETY ADVISORY NOTICE" dated December 12, 2011 to the affected customer. The letter identifies the product, problem, and actions to be taken by the customer. The firm prepared a modification, which will be provided to the affected customer under Update Instruction XP055/11/S. The firm will resolve this potential issue by preventively exchanging all bolts for the fixation of the X-ray tube holder of the concerned AXIOM Aristos FX Plus unit.
Quantity
1
Distribution
Product distributed in one hospital in California.
Lot, serial or model codes
Serial number 1273
510(k) numbers
["K061054"]
PMA numbers
not on file
Product code
MQB
Firm FEI
2240869
Firm city
Malvern
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register