Boston Scientific Corporation, Neurovascular, 47900 Bayside Parkway, Fremont, CA 94538: Boston Scientific brand Matrix2 360 Standard SR Coils. Intended for embolization of certain intracranial aneurysms and for arterial and venous embolizations in the peripheral vasculature.

Recall number
Z-0860-2011
Recalling firm
Boston Scientific Corporation, Neurovascular, 47900 Bayside Parkway, Fremont, CA 94538
Product
Boston Scientific brand Matrix2 360 Standard SR Coils. Intended for embolization of certain intracranial aneurysms and for arterial and venous embolizations in the peripheral vasculature.
Status
Terminated
Initiated
2010-10-19
Posted
2011-01-05
Terminated
2012-01-27
Reason
The product appears to have premature degradation of the PGLA coating on the Matrix Coil.
Root cause
Process control
Action
US consignees were notified by letter sent by Federal Express on October 19, 2010. Customers out of the United States were notified by Federal Express if possible and by personal delivery by locally assigned sales representatives as needed.
Quantity
77,364 total units
Distribution
Worldwide distribution. USA, Argentina, Austria, Australia, Belgium, Bahrain, Brazil, Canada, Switzerland, Chile, China, Colombia, Serbia, Czech Republic, Germany, Denmark, Spain, Finland, France, Great Britain, Greece, Hong Kong, Hungary, Ireland, India, Israel, Italy, Jordan, Japan, South Korea, Lithuania, Latvia, Macau, Mexico, Malaysia, Netherlands, Norway, New Zealand, Panama, Philippines, Poland, Portugal, Qatar, Romania, Russian Federation, Saudi Arabia, Sweden, Singapore, Slovakia, Thailand, Taiwan, Turkey, Ukraine, Vietnam, and Yemen.
Lot, serial or model codes
Material Product Description Catalog # Batch numbers Expiry Date    11322013,11341317,11395760,11407118,   11484317,11500845,11538872,11580676,   11586996,11591704,11676172,11786207,   11800707,11843132,11889308,11889310,   11924294, 11946182, 11975161, 11975197,   Matrix2¿ 360 Standard SR Coils 12000345,12119829,12127493,12156626, Nov2010 through  M003492103SRO 492103-SR 12176110, 12194735, 12214709, 12214800,   10mm x 30cm 12214853, 12229661, 12291690, 12299408, Jun2013   12315367,12330591,12409146,12418207,   12439180,12439189,12514594,12669964,   12670044,12703195,12791152,12792881,   12863906,12909368,12913888,12921982,   13165100,13182071,13602875   11296519,11329483,11422711,11477789,   11606464,11711221,11712103,11786210,   11908091,11924295,11940396,11975180,   11996498,12009649,12112236,12119825,   Matrix2¿ 360 Standard SR Coils 12141857,12156617,12214707,12214715, Oct2010 through  M003492123SRO 492123-SR 12214861,12291695,12295742,12314469,   12mm x 30cm 12392928,12401322,12458011,12462033, Jun2013   12519869, 12523479, 12669970, 12810344,   12861691,12912140,12920952,12921980,   12921996,12982276,13164798,13165084,   13603462   Matrix2¿ 360 Standard SR Coils 11286172,11327021,11552550,11756832, Oct2010 through  M003492308SRO 492308-SR 11792451,12082838,12194728,12494078,   3mm x8cm 12947522 Aug2012   11310925,11340517,11513571,11598083,   11748422,11820289,11850852,11924296,   11940398,11975077,12007427,12026497,   Matrix2¿ 360 Standard SR Coils 12026741,12074031,12074306,12110477, Oct2010 through  M003492408SRO 492408-SR 12130818,12142084,12156609,12176095,   4mm x8cm 12176129,12214851,12304396,12418211, Dec2012   12434605, 12439350, 12579013, 12702894,   12721623,12792899,12810370,12917824,   12947545, 13164799   11289583,11341604,11500842,11622022,  M003492510SRO Matrix2¿ 360 Standard SR Coils 492510-SR 11730241, 11786201, 11820321, 12194490, Oct2010 through   5mm x 10cm 12214705, 12434606, 12450913, 12465570, Aug2012   12922017   11299199,11318111,11389194,11500843,   11550742,11606466,11697460,11711220,   11748398,11838629,11889318,11916974,   11958443,11958463,11984850,12043111,   12061948,12082851,12086034,12142091,  M003492515SRO Matrix2¿ 360 Standard SR Coils 492515-SR 12150131,12176136,12188226,12194738, Oct2010 through   5mmx15cm 12214194,12225443,12229718,12330640, Aug2013   12361044,12378759,12383763,12431052,   12439197, 12636726, 12656660, 12792894,   12805887,12861538,12861680,12899138,   12921993,12922024,12929260,12982324,   13164797,13182067,13197169,13740709   Matrix2¿ 360 Standard SR Coils 11300189,11697441,11748423,12102562, Oct2010 through  M003492610SRO 492610-SR 12194744,12650932,13109926,13201013,   6mm x 10cm 13220484 Dec2012   11281622,11311916,11314681,11491189,   11521070, 11591707, 11676160, 11730225,   11766704,11838625,11934378,11958468,   11975166,11975191,11992203,11992212,   12009577,12009653,12082842,12130810,   Matrix2¿ 360 Standard SR Coils 12142036,12161320,12168019,12176126, Oct2010 through  M003492615SRO 492615-SR 12214841,12229589,12300888,12330615,   6mmx15cm 12379816, 12388035,12409150,12427992, Dec2012   12439353, 12519870, 12549504, 12549518,   12636727, 12670053, 12721624, 12748411,   12791149, 12792892, 12861690, 12889253,   12912271,12922000,12947541,12982277,   13016751,13165091   Matrix2¿ 360 Standard SR Coils 11311649,11552534,11697440,11697450, Oct2010 through  M003492715SRO 7mm x 15cm 492715-SR 11766703,12176124,12229551,12372622, Dec2012   12494180,12698228,12922016,13201014 11325958,11330877,11448223,11516241,   11591599, 11640896, 11676142, 11676149,   11766671, 11786209, 11795499, 11820176,   11868816,11925596,11975072,11975182,   11992202,11992218,12009562,12009570,  M003492720SRO Matrix2¿ 360 Standard SR Coils 492720-SR 12051397,12082854,12142030,12188227, Nov2010 through   7mm x20cm 12194711,12194747,12214845,12225440, Jul2013   12291691,12299412,12314471,12324046,   12355111,12401321,12409152,12418208,   12427990,12636754,12810346,12841940,   12861722, 12917822, 12922005, 12947520,   12947528, 13165089, 13705204   11303726, 11344541, 11491183, 11597978,   11697453,11730059,11766677,11800708,  M003492820SRO Matrix2¿ 360 Standard SR Coils 492820-SR 11836538,11838623,11958239,12043226, Oct2010 through   8mm x20cm 12130809,12194736,12260123,12388030, Jun2013   12669969, 12791148, 12792880, 12922002,   13201015, 13202909, 13606213   11310595,11410936,11470446,11552546,   11606467,11658788,11712109,11766665,   11800187,11838294,11869782,11925588,   11975179,11992216,12009571,12043110,   Matrix2¿ 360 Standard SR Coils 12061953, 12069212, 12074311, 12119903, Oct2010 through  M003492830SRO 492830-SR 12150137,12188230,12194489,12194704,   8mm x30cm 12229669, 12291686, 12296707, 12314472, Dec2012   12330585,12356257,12361056,12636753,   12669979, 12703142, 12753169, 12783354,   12792913, 12895634, 12899136, 12912676,   12922009,13165096  .,tl003492920SRO Matrix2¿ 360 Standard SR Coils 492920-SR 11318113,11441724,11542877,11711222, Nov2010 through   9mm x20cm 11838626,12150132,12579018,13614169 Jun2013   11290201,11318378,11403946,11488185,   11520165,11598081,11658797,11748390,   11889313,11940587,11952858,12009643,   12026498,12061945,12069211,12074037,   Matrix2¿ 360 Standard SR Coils 12130804,12176121,12194485,12194498, Oct2010 through  M003492930SRO 492930-SR 12194734,12229597,12295451,12296708,   9mm x 30cm 12325703,12355110,12549418,12549503, Aug2013   12564285,12579005,12589397,12656661,   12669971,12792539,12792897,12853506,   12861697,12891631,12922012,12982278,   13182063, 13204832, 13740457 .
510(k) numbers
["K012985","K031168"]
PMA numbers
not on file
Product code
HGC
Firm FEI
2939204
Firm city
Fremont
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register