FDA Medical Device Recalls (openFDA)
Abbott Laboratories MPG: ARCHITECT B12 Reagent, list 06C09-22 (4 x 100 tests), 06C09-27 (1 x 100 tests); an in vitro diagnostic kit consisting of microparticles, conjugate, assay diluent, pre-treatment 1, pre-treatment 2, and pre-treatment 3; Abbott Laboratories, Abbott Park, IL 60064
- Recall number
- Z-0860-06
- Recalling firm
- Abbott Laboratories MPG
- Product
- ARCHITECT B12 Reagent, list 06C09-22 (4 x 100 tests), 06C09-27 (1 x 100 tests); an in vitro diagnostic kit consisting of microparticles, conjugate, assay diluent, pre-treatment 1, pre-treatment 2, and pre-treatment 3; Abbott Laboratories, Abbott Park, IL 60064
- Status
- Terminated
- Initiated
- 2006-03-08
- Posted
- 2006-05-11
- Terminated
- 2006-07-31
- Reason
- The 30-day onboard storage information is not included in the barcode labels for these two lots. As a result, the ARCHITECT system software is unable to track how long a reagent kit has been stored onboard the ARCHITECT instrument.
- Root cause
- Other
- Action
- Abbott affiliates in Canada, Germany and Australia were e-mailed copies of the recall letter on 3/8/06 for translation where needed and dissemination to their accounts receiving the suspect lots. The end users were advised that the 30-day onboard storage information is not included in the barcode for the two lots. As a result the ARCHITECT system software is unable to track how long the reagent kit has been stored onboard. The accounts were instructed to manually track the onboard storage time for the two lots, and discard the kits after 30 days of onboard storage.
- Quantity
- 135 kits
- Distribution
- Canada, Germany and Australia
- Lot, serial or model codes
- lot numbers 33079M200, 33079M201
- 510(k) numbers
- ["K984108"]
- PMA numbers
- not on file
- Product code
- LIG
- Firm FEI
- 1415939
- Firm city
- Abbott Park
- Firm state
- IL
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28