FDA Medical Device Recalls (openFDA)
CareFusion 203, Inc.: Cardinal Health EnVe Ventilator Cat # 19250-001, Manufactured by Cardinal Health, Minneapolis, MN 55447-1341. Intended to provide continuous or intermittent ventilatory support for the care of individuals who require mechanical ventilation.
- Recall number
- Z-0857-2010
- Recalling firm
- CareFusion 203, Inc.
- Product
- Cardinal Health EnVe Ventilator Cat # 19250-001, Manufactured by Cardinal Health, Minneapolis, MN 55447-1341. Intended to provide continuous or intermittent ventilatory support for the care of individuals who require mechanical ventilation.
- Status
- Terminated
- Initiated
- 2010-01-08
- Posted
- 2010-02-25
- Terminated
- 2011-12-17
- Reason
- CareFusion is recalling the EnVe Ventilator because the EnVe Ventilator does not fully meet applicable standards for immunity to electro-Magnetic Interference (EMI) when tested in the laboratory environment. CareFusion has identified certain device malfunctions that may affect device performance and patient safety in the intended use environment.
- Root cause
- Component design/selection
- Action
- CareFusion issued an "Urgent Medical Device Recall notification" letter dated January 07, 2010. The letter was addressed to the Director of Respiratory Care describing the problem and the product. Consignees were told that a Service Representative will contact them to arrange for the prompt return of the affected product for correction. For further information, contact CareFusion Technical Support at 1-800-231-2446, Option 1.
- Quantity
- 9
- Distribution
- MN
- Lot, serial or model codes
- Serial number# 33600, 33604, 33619, 35063, 35068, 35071, 35072, 35073, 35074.
- 510(k) numbers
- ["K070594"]
- PMA numbers
- not on file
- Product code
- CBK
- Firm FEI
- 3002617992
- Firm city
- Minneapolis
- Firm state
- MN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28