Skip to the content

FDA Medical Device Recalls (openFDA)

CareFusion 203, Inc.: Cardinal Health EnVe Ventilator Cat # 19250-001, Manufactured by Cardinal Health, Minneapolis, MN 55447-1341. Intended to provide continuous or intermittent ventilatory support for the care of individuals who require mechanical ventilation.

Recall number
Z-0857-2010
Recalling firm
CareFusion 203, Inc.
Product
Cardinal Health EnVe Ventilator Cat # 19250-001, Manufactured by Cardinal Health, Minneapolis, MN 55447-1341. Intended to provide continuous or intermittent ventilatory support for the care of individuals who require mechanical ventilation.
Status
Terminated
Initiated
2010-01-08
Posted
2010-02-25
Terminated
2011-12-17
Reason
CareFusion is recalling the EnVe Ventilator because the EnVe Ventilator does not fully meet applicable standards for immunity to electro-Magnetic Interference (EMI) when tested in the laboratory environment. CareFusion has identified certain device malfunctions that may affect device performance and patient safety in the intended use environment.
Root cause
Component design/selection
Action
CareFusion issued an "Urgent Medical Device Recall notification" letter dated January 07, 2010. The letter was addressed to the Director of Respiratory Care describing the problem and the product. Consignees were told that a Service Representative will contact them to arrange for the prompt return of the affected product for correction. For further information, contact CareFusion Technical Support at 1-800-231-2446, Option 1.
Quantity
9
Distribution
MN
Lot, serial or model codes
Serial number# 33600, 33604, 33619, 35063, 35068, 35071, 35072, 35073, 35074.
510(k) numbers
["K070594"]
PMA numbers
not on file
Product code
CBK
Firm FEI
3002617992
Firm city
Minneapolis
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register