Boston Scientific Corporation, Neurovascular, 47900 Bayside Parkway, Fremont, CA 94538: Boston Scientific brand Matrix 2D Soft SR Coils. Intended for embolization of certain intracranial aneurysms and for arterial and venous embolizations in the peripheral vasculature.

Recall number
Z-0856-2011
Recalling firm
Boston Scientific Corporation, Neurovascular, 47900 Bayside Parkway, Fremont, CA 94538
Product
Boston Scientific brand Matrix 2D Soft SR Coils. Intended for embolization of certain intracranial aneurysms and for arterial and venous embolizations in the peripheral vasculature.
Status
Terminated
Initiated
2010-10-19
Posted
2011-01-05
Terminated
2012-01-27
Reason
The product appears to have premature degradation of the PGLA coating on the Matrix Coil.
Root cause
Process control
Action
US consignees were notified by letter sent by Federal Express on October 19, 2010. Customers out of the United States were notified by Federal Express if possible and by personal delivery by locally assigned sales representatives as needed.
Quantity
77,364 total units
Distribution
Worldwide distribution. USA, Argentina, Austria, Australia, Belgium, Bahrain, Brazil, Canada, Switzerland, Chile, China, Colombia, Serbia, Czech Republic, Germany, Denmark, Spain, Finland, France, Great Britain, Greece, Hong Kong, Hungary, Ireland, India, Israel, Italy, Jordan, Japan, South Korea, Lithuania, Latvia, Macau, Mexico, Malaysia, Netherlands, Norway, New Zealand, Panama, Philippines, Poland, Portugal, Qatar, Romania, Russian Federation, Saudi Arabia, Sweden, Singapore, Slovakia, Thailand, Taiwan, Turkey, Ukraine, Vietnam, and Yemen.
Lot, serial or model codes
Material Product Description Catalog # Batch Numbers Expiry Date   M003471304SRO Matrix2¿ 2D Soft SR Coils 3mm 471304-SR 11391991,11697446,11721433,11775115, Dec2010 through   x4cm 12579020 Mar2012  M003471306SRO Matrix2¿ 2D Soft SR Coils 3mm 471306-SR 11498868,11775116,12194493,12385403, Jan2011 through   x6cm 12520966, 12861735 Jul2012  M003471308SRO Matrix2¿ 2D Soft SR Coils 3mm 471308-SR 11481348,11792449,11820171,12401292, Jan2011 through   x8cm 12514514, 12792920 Jun2012  M003471310SRO Matrix2¿ 2D Soft SR Coils 3mm 471310-SR 11342011,11634830,12390151,12439220, Nov2010 through   x 10cm 12982264 Sep2012  M003471406SRO Matrix2¿ 2D Soft SR Coils 4mm 471406-SR 11502704,11641203,12081496,12474445, Jan2011 through   x6cm 12795286 Jun2012   11622088,11718835,11869785,11925599,   11940238,11985921,11992220,12156354,  M003471408SRO Matrix2¿ 2D Soft SR Coils 4mm 471408-SR 12156603,12176007,12176017,12194725, Mar2011 through   x8cm 12295449,12356324,12356325,12514591, Dec2012   12636738,12748405,12748465,12845925,   12891630,12906595,13200124   Matrix2¿ 2D Soft SR Coils 4mm 11310206,11552557,11766672,12119904, Oct2010 through  M003471410SRO 471410-SR 12161473,12409142,12439231,12791147,   x 10cm 12792538 Jun2012  M003471508SRO Matrix2¿ 2D Soft SR Coils 5mm 471508-SR 11622083,12176098,12229542,12721621, Mar2011 through   x8cm 13020445 Sep2012   11800192,11869788,11925606,11940381,   11963592,11996801,12155987,12156601,  M003471510SRO Matrix2¿ 2D Soft SR Coils 5mm 471510-SR 12176008,12194708,12314468,12330568, May2011 through   x 10cm 12368755,12409141,12514589,12636752, Jun2013   12669963,12670049,12703140,12753166,   12792908,12861687,12981682,13603463  M003471515SRO Matrix2¿ 2D Soft SR Coils 5mm 471515-SR 11384554,11641206,12176118,12182654, Dec2010 through   x 15cm 12246986,12649894,13020067 Sep2012   11384857,11748388,11925177,11940389,  M003471610SRO Matrix2¿ 2D Soft SR Coils 6mm 471610-SR 11967352,12112307,12155988,12176015, Dec2010 through   x 10cm 12330561,12356327,12374220,12847321, Aug2012   12947525,12971457  M003471615SRO Matrix2¿ 2D Soft SR Coils 6mm 471615-SR 11532410,11800196 Feb2011 through   x 15cm May2011  1003471620SRO Matrix2¿ 2D Soft SR Coils 6mm 471620-SR 11373732,11634833,12261958,12649895, Dec2010 through   x20cm 12792911 Jun2012  M003471710SRO Matrix2¿ 2D Soft SR Coils 7mm 471710-SR 12499178,13108359 Feb2012 through   x 10cm Oct2012  M003471715SRO Matrix2¿ 2D Soft SR Coils 7mm 471715-SR 11820280,12229546,12249163 May2011 through   x 15cm Nov2011  M003471720SRO Matrix2¿ 2D Soft SR Coils 7mm 471720-SR 11622084,11938494,12246985,12579366 Mar2011 through   x20cm Mar2012  M003471820SRO Matrix2¿ 2D Soft SR Coils 8mm 471820-SR 11622082,11766666 Mar2011 through   x20cm May2011  M003471830SRO Matrix2¿ 2D Soft SR Coils 8mm 471830-SR 11697466,12229584 Apr2011 through   x30cm Nov2011
510(k) numbers
["K012985","K031168"]
PMA numbers
not on file
Product code
HGC
Firm FEI
2939204
Firm city
Fremont
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register