Skip to the content

FDA Medical Device Recalls (openFDA)

Beckman Coulter Inc.: UniCel DxI 800 and 600 Access Immunoassay Systems, UniCel DxC 880i, 860i, 660i, and 680i SYNCHRON Access Clinical Systems with Dxl System Software versions 4.2 or 4.2.1. In vitro diagnostic device used for the quantitative, semi-quantitive or qualitative determination of various analyte concentrations found in human body fluids.

Recall number
Z-0856-2010
Recalling firm
Beckman Coulter Inc.
Product
UniCel DxI 800 and 600 Access Immunoassay Systems, UniCel DxC 880i, 860i, 660i, and 680i SYNCHRON Access Clinical Systems with Dxl System Software versions 4.2 or 4.2.1. In vitro diagnostic device used for the quantitative, semi-quantitive or qualitative determination of various analyte concentrations found in human body fluids.
Status
Terminated
Initiated
2009-10-28
Posted
2010-02-25
Terminated
2012-05-07
Reason
This recall was initiated because of software issue was identified in System Software 4.2 and 4.2.1, which may result in the incorrect handling of a wash arm dispense plate motion failure. Instead of generating a red warning event and stopping sample processing, the system will post a yellow caution event and continue to operate with the dispense plate in an elevated position until the instrument
Root cause
Software design
Action
Beckman Coulter issued an "Urgent: Product Corrective Action" letter dated October 28, 2009. Customers were notified of the affected product and how the issue will be resolved. For further information, contact Beckman Coulter Inc. Technical Support at 1-800-854-3633.
Quantity
377 Instruments
Distribution
Worldwide Distribution -- United States and Canada.
Lot, serial or model codes
DxI software versions: 4.2 and 4.2.1. For labeling see UniCel DxI Operator's Guide. Part Number: 387262G
510(k) numbers
["K023764","K060256"]
PMA numbers
not on file
Product code
JJE
Firm FEI
2050012
Firm city
Brea
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register