Skip to the content

FDA Medical Device Recalls (openFDA)

Accuray Inc: RoboCouch Patient Support System, a component of the CyberKnife Robotic Radiosurgery System. Model number 025007, manufactured by Accuray Inc., Sunnyvale, CA The CyberKnife is indicated for treatment planning and imageguided stereotactic radiosurgery and precision radiotherapy of lesions, tumors and conditions anywhere in the body when radiation treatment is indicated. The RoboCouch Patient Support System is intended for use in the support and positioning of a patient during Radiosurgery and radiotherapy procedures and other medical procedures when precise positioning is required.

Recall number
Z-0856-2009
Recalling firm
Accuray Inc
Product
RoboCouch Patient Support System, a component of the CyberKnife Robotic Radiosurgery System. Model number 025007, manufactured by Accuray Inc., Sunnyvale, CA The CyberKnife is indicated for treatment planning and imageguided stereotactic radiosurgery and precision radiotherapy of lesions, tumors and conditions anywhere in the body when radiation treatment is indicated. The RoboCouch Patient Support System is intended for use in the support and positioning of a patient during Radiosurgery and radiotherapy procedures and other medical procedures when precise positioning is required.
Status
Terminated
Initiated
2008-10-30
Posted
2009-01-14
Terminated
2009-09-10
Reason
Product may not be tensioned properly, potentially causing unexpected rotation or descent.
Root cause
Process control
Action
All affected sites were notified of the potential problem via Customer Advisory Notice dated 30 October 2008. The letter states required actions that the customer should follow at the beginning of each treatment day. The tension of the belt drive and the support bolts in question will be inspected by Accuray Field Service personnel at the affected facilities. In addition, the preventive maintenance procedure will be updated to include regular inspection and replacement of the belt-drive. Any questions or concerns regarding this issue should be directed to Accuray Customer Support at 1-877-668-8667 or 1-408-716-4700 or customersupport@accuray.com.
Quantity
13 units
Distribution
Worldwide Distribution --- including USA and countries of France and Turkey.
Lot, serial or model codes
Model number 025007, system numbers C0158, C0078, C0141, C0160, C0128, C0178, C0179, C0152, C0110, C0113, C0130, C0060, and C0125.
510(k) numbers
["K042146"]
PMA numbers
not on file
Product code
JAI
Firm FEI
1000123585
Firm city
Sunnyvale
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register