Skip to the content

FDA Medical Device Recalls (openFDA)

Orthofix, Inc.: Orthofix Firebird Spinal Fixation System Instrumentation Modular Screw Driver, Product Usage: The Modular Screw Driver is used for inserting a modular bone screw into the pedicle during a spinal fixation procedure. It is a reusable instrument.

Recall number
Z-0855-2014
Recalling firm
Orthofix, Inc.
Product
Orthofix Firebird Spinal Fixation System Instrumentation Modular Screw Driver, Product Usage: The Modular Screw Driver is used for inserting a modular bone screw into the pedicle during a spinal fixation procedure. It is a reusable instrument.
Status
Terminated
Initiated
2013-11-26
Posted
2014-01-28
Terminated
2014-12-09
Reason
Orthofix received 6 complaints which resulted in a reportable events due to extended surgical times greater than 30 minutes for the Modular Screw Driver (PN 52-1332). The complaints alleged that the screw driver's collet would malfunction resulting in the surgeon being unable to use the Modular Screw Driver to effectively place Modular Screws, which may result in a delay of surgery.
Root cause
Nonconforming Material/Component
Action
Orthofix sent an Urgent Medical Device Recall Notification letter dated November 26, 2013 by certified mail to distributors. The letter identified the affected product, problem and actions to be taken. The letter instructed distributors to immediately cease any further distribution, return affected products to Orthofix at no cost and complete the attached reply form. For questions contact your local Orthofix representative.
Quantity
443
Distribution
Worldwide Distribution - US Nationwide including (Puerto Rico) and the countries of: Spain, Germany, Australia, Peru, Mexico, Colombia, South Africa, and Israel.
Lot, serial or model codes
Production Identification Numbers: Part Number 52-1332, All Lot Numbers.
510(k) numbers
["K081684"]
PMA numbers
not on file
Product code
KWP
Firm FEI
2183449
Firm city
Lewisville
Firm state
TX
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register