Varian Medical Systems Oncology Systems: Eclipse 8.5 (build 8.2.2X) systems with distributed Calculation Framework (DCF) 8.2.22, model number H48, manufactured by Varian Medical Systems, Inc., Palo Alto, CA. Indicated for planning proton treatment of neoplasms of the eye.

Recall number
Z-0855-2009
Recalling firm
Varian Medical Systems Oncology Systems
Product
Eclipse 8.5 (build 8.2.2X) systems with distributed Calculation Framework (DCF) 8.2.22, model number H48, manufactured by Varian Medical Systems, Inc., Palo Alto, CA. Indicated for planning proton treatment of neoplasms of the eye.
Status
Terminated
Initiated
2008-10-02
Posted
2009-01-15
Terminated
2010-11-27
Reason
When using the multiple static segment option, the device may alter dose delivery.
Root cause
Software design
Action
An Urgent Medical Device Correction Letter was distributed to all affected users, with a description of the problem and user corrective action steps on October 2, 2008. The Letter will also be distributed to the Varian Sales, Marketing, and Service organizations, informing them of the issue. The firm recommends that customers DO NOT USE the Multiple Static Segment (MSS) option in Varian Leaf Motion Calculator version 8.2.22. The firm also recommends setting the default delivery method for Varian Leaf Motion Calculator to Sliding Window. When using Eclipse 8.5, always follow the Safety Instructions found in chapter 2 of "Eclipse External Beam Planning Instruction for Use", P/N B501357R01A, March 2008. If customers believe a patient treatment has been affected by this issue, they should notify a Varian Service or Support person as soon as possible.
Quantity
64 units
Distribution
Worldwide Distribution --- including USA and countries of Switzerland and the Netherlands.
Lot, serial or model codes
Model number H48, serial numbers H480016, H480051, H480052, H480053, H480054, H480055, H480107, H480133, H480293, H480346, H480347, H480472, H480652, H480920, H481467, H481510, H481639, H481724, H481807, H481815, H481816, H481912, H481913, H481914, H481915, H482194, H482195, H482289, H482793, H482807, H483204, H483407, H484367, H484635, H484636, H484637, H484638, H484781, H484864, H485100, H485125, H485244, H485346, H485466, H485729, H486858, H486859, H486860, H486861, H486946, H486947, H486948, H487190, H487212, H487414, H487580, H487581, H487623, H487640, H487641, H487642, H487700, and H487757.
510(k) numbers
["K071873"]
PMA numbers
not on file
Product code
MUJ
Firm FEI
2916710
Firm city
Palo Alto
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register