Skip to the content

FDA Medical Device Recalls (openFDA)

Abbott Laboratories MPG: ARCHITECT STAT Troponin-I Reagent; list numbers 2K41-20 (400 test kit), 2K41-25 (100 test kit) and 2K41-30 (2000 test kit); each kit contains bottles of microparticles, conjugate and assay diluent; Abbott Laboratories, Abbott Park, IL 60064

Recall number
Z-0855-2007
Recalling firm
Abbott Laboratories MPG
Product
ARCHITECT STAT Troponin-I Reagent; list numbers 2K41-20 (400 test kit), 2K41-25 (100 test kit) and 2K41-30 (2000 test kit); each kit contains bottles of microparticles, conjugate and assay diluent; Abbott Laboratories, Abbott Park, IL 60064
Status
Terminated
Initiated
2007-03-08
Posted
2007-06-14
Terminated
2008-07-16
Reason
False Negative Results : The analytical sensitivity claim of less than or equal to 0.01 ng/mL (ug/L) might not be met for all lots of Architect STAT Troponin-I reagents, thus false negative results may occur with negative results less than 0.1 ng/mL
Root cause
Other
Action
Abbott sent a Product Correction letter on 3/8/07 to all Troponin-I customers who received reagent lots in the last 12 months. The letters informed the accounts that the analytical sensitivity claim of = 0.01 ng/mL might not be met for all lots of Architect STAT Troponin-I reagents. The accounts were directed to evaluate the information the information provided in the letter against the reference range used in their laboratory to determine if any action needs to be taken; to follow their facility's policies and procedures for communicating with the heath care providers they serve; to forward a copy of the letter to any laboratory to whom they may have forwarded the reagent; and to retain a copy of the letter for their records. Any questions were directed to Abbott Customer Support at 1-877-4ABBOTT in the U.S. A follow-up press release was issued on 5/25/07 and a follow-up letter was issued on 5/29/07, providing updated instructions for laboratories to follow when using the product.
Quantity
9,438 kits
Distribution
Worldwide: USA including Puerto Rico, and internationally to Canada, Brazil, Germany, Japan, Taiwan, Hong Kong, Thailand, South Korea, Australia, New Zealand, Mexico, Colombia, Chile, Argentina and Trindad & Tobago.
Lot, serial or model codes
Lot numbers 11455C006, 11605UN06, 12072UN06, 14818UN06, 13037UN06, 12990UN06, 12982UN06, 11456C006, 11609UN06, 14817UN06, 14339UN06, 13590UN06, 12986UN06, 14350UN06, 13122UN06, 12978UN06, 11370C006
510(k) numbers
["K041192"]
PMA numbers
not on file
Product code
MMI
Firm FEI
1415939
Firm city
Abbott Park
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register