FDA Medical Device Recalls (openFDA)
Zimmer, Inc.: Persona Cemented Tibial Broach Inserter/Extractor. The Persona Cemented Tibial Broach Inserter/Extractor Handle is a Use sterilizable instrument intended for use In multiple TKA procedures. The Persona Surgical Technique indicates that the Persona Cemented Tibial Broach Inserter/Extractor Handle is intended to broach the proximal tibia in preparation for the full seating of the tibial implant.
- Recall number
- Z-0853-2015
- Recalling firm
- Zimmer, Inc.
- Product
- Persona Cemented Tibial Broach Inserter/Extractor. The Persona Cemented Tibial Broach Inserter/Extractor Handle is a Use sterilizable instrument intended for use In multiple TKA procedures. The Persona Surgical Technique indicates that the Persona Cemented Tibial Broach Inserter/Extractor Handle is intended to broach the proximal tibia in preparation for the full seating of the tibial implant.
- Status
- Terminated
- Initiated
- 2014-12-04
- Posted
- 2014-12-23
- Terminated
- 2015-05-13
- Reason
- The affected lot was supplied by contract manufacturer and are potentially subject to disassociation of the magnet cover. Zimmer received one complaint indicating the magnet cover of a Persona Cemented Tibial Broach Inserter/Extractor Handle lot 62085420 was missing after broaching the tibia. Postoperative radiographs were inspected and the magnet cover was not identified. The investigation indi
- Root cause
- Process design
- Action
- Notifications will be sent to the affected distributors and hospital risk managers with instructions for returning the affected product the week of December 8, 2014. Recalled product will be quarantined and returned to Zimmer Sales Representatives to be destroyed. The public contact for the recall is: Consumer Relations Call Center Zimmer, Inc. P O Box 708 Warsaw, IN 46581-0708 consumer@zimmer.com Phone: 1-800-447-5633 Fax: 574-371-8755
- Quantity
- 83 units
- Distribution
- Distributed in the states of MN; NJ; NC; PA; AL; MI; OR; FL; OH; CO; IL and CA, and the countries of Australia, India, UAE, Germany, and Italy.
- Lot, serial or model codes
- Item: 42-5399-023-00 Lots: 62072307 62072565 62085420 62192754
- 510(k) numbers
- ["K113369"]
- PMA numbers
- not on file
- Product code
- JWH
- Firm FEI
- 1000220733
- Firm city
- Warsaw
- Firm state
- IN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28