Skip to the content

FDA Medical Device Recalls (openFDA)

Zimmer, Inc.: Persona Cemented Tibial Broach Inserter/Extractor. The Persona Cemented Tibial Broach Inserter/Extractor Handle is a Use sterilizable instrument intended for use In multiple TKA procedures. The Persona Surgical Technique indicates that the Persona Cemented Tibial Broach Inserter/Extractor Handle is intended to broach the proximal tibia in preparation for the full seating of the tibial implant.

Recall number
Z-0853-2015
Recalling firm
Zimmer, Inc.
Product
Persona Cemented Tibial Broach Inserter/Extractor. The Persona Cemented Tibial Broach Inserter/Extractor Handle is a Use sterilizable instrument intended for use In multiple TKA procedures. The Persona Surgical Technique indicates that the Persona Cemented Tibial Broach Inserter/Extractor Handle is intended to broach the proximal tibia in preparation for the full seating of the tibial implant.
Status
Terminated
Initiated
2014-12-04
Posted
2014-12-23
Terminated
2015-05-13
Reason
The affected lot was supplied by contract manufacturer and are potentially subject to disassociation of the magnet cover. Zimmer received one complaint indicating the magnet cover of a Persona Cemented Tibial Broach Inserter/Extractor Handle lot 62085420 was missing after broaching the tibia. Postoperative radiographs were inspected and the magnet cover was not identified. The investigation indi
Root cause
Process design
Action
Notifications will be sent to the affected distributors and hospital risk managers with instructions for returning the affected product the week of December 8, 2014. Recalled product will be quarantined and returned to Zimmer Sales Representatives to be destroyed. The public contact for the recall is: Consumer Relations Call Center Zimmer, Inc. P O Box 708 Warsaw, IN 46581-0708 consumer@zimmer.com Phone: 1-800-447-5633 Fax: 574-371-8755
Quantity
83 units
Distribution
Distributed in the states of MN; NJ; NC; PA; AL; MI; OR; FL; OH; CO; IL and CA, and the countries of Australia, India, UAE, Germany, and Italy.
Lot, serial or model codes
Item: 42-5399-023-00 Lots: 62072307 62072565 62085420 62192754
510(k) numbers
["K113369"]
PMA numbers
not on file
Product code
JWH
Firm FEI
1000220733
Firm city
Warsaw
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register