Skip to the content

FDA Medical Device Recalls (openFDA)

Baxter Healthcare Corp.: Baxter PosiFlow Access Device for IV Access; product code 2N9050; a single use, sterile device with PosiFlow Positive Displacement Feature and Male Luer Lock Adapter, 0.06 mL; Baxter Healthcare Corp., Deerfield, IL 60015 USA; the PosiFlow Access Device is manufactured for Baxter Healthcare Corporation by Becton Dickinson

Recall number
Z-0853-05
Recalling firm
Baxter Healthcare Corp.
Product
Baxter PosiFlow Access Device for IV Access; product code 2N9050; a single use, sterile device with PosiFlow Positive Displacement Feature and Male Luer Lock Adapter, 0.06 mL; Baxter Healthcare Corp., Deerfield, IL 60015 USA; the PosiFlow Access Device is manufactured for Baxter Healthcare Corporation by Becton Dickinson
Status
Terminated
Initiated
2005-03-03
Posted
2005-06-01
Terminated
2006-05-11
Reason
Inadequate directions for use resulting in fluid leaks, valves sticking in the open position, and flow output restriction.
Root cause
Other
Action
Urgent Device Correction letters dated 3/3/05 were sent to the direct accounts via first class mail, to the attention of the Director of Nursing. The accounts were informed of the field reports received concerning the PosiFlow Access Device and IV sets containing the PosiFlow device and the changes made to the directions for use and in-service training materials to correct the problems. The accounts were requested to complete and fax back to Baxter at 847-270-5457 the enclosed customer reply form, acknowledging the receipt and dissemination of the information to their staff. Wholesalers were requested to notify their customers. Any questions were directed to the Center for One Baxter at 1-800-422-9837.
Quantity
not known
Distribution
Nationwide
Lot, serial or model codes
product code 2N9050, all lots
510(k) numbers
["K964918"]
PMA numbers
not on file
Product code
FPA
Firm FEI
1416980
Firm city
Round Lake
Firm state
IL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register