Skip to the content

FDA Medical Device Recalls (openFDA)

Stryker Spine: AccuLIF PL Tubing Set and AccuLIF TL Tubing Set AccuLIF is an interbody device system that uses hydraulic pressure to expand the implant in-situ. The purpose of the AccuLIF PL Tubing Set (cat#900301) and AccuLIF TL Tubing Set (cat#905103) is to deliver pressurized saline from a syringe to the AccuLIF implant for in-situ expansion. The Tubing Set attaches to the AccuLIF Insertion handle at two district locations: the "junction block" and the "manifold." The PL and TL Tubing Sets are Class 2 instruments not intended to sustain or support life.

Recall number
Z-0852-2016
Recalling firm
Stryker Spine
Product
AccuLIF PL Tubing Set and AccuLIF TL Tubing Set AccuLIF is an interbody device system that uses hydraulic pressure to expand the implant in-situ. The purpose of the AccuLIF PL Tubing Set (cat#900301) and AccuLIF TL Tubing Set (cat#905103) is to deliver pressurized saline from a syringe to the AccuLIF implant for in-situ expansion. The Tubing Set attaches to the AccuLIF Insertion handle at two district locations: the "junction block" and the "manifold." The PL and TL Tubing Sets are Class 2 instruments not intended to sustain or support life.
Status
Terminated
Initiated
2015-12-16
Posted
not on file
Terminated
2016-08-01
Reason
There were two potential interference conditions identified with the way the tubing set attaches to the inserter.
Root cause
Device Design
Action
Stryker Spine sent an Urgent Product Removal letter dated December 17, 2015 to affected customers via Fed Ex priority overnight. The letter identified the affected product, problem and actions to be taken. Customers are asked to follow instructions included in the letter. The customer response form should be completed and sent back to email via Spine-RegulatoryActions@Stryker.com or Fax 855-632-9049. For questions call (201)749-8389.
Quantity
413 units
Distribution
US Nationwide Distribution
Lot, serial or model codes
Catalog number 900301, Lot #01131543, 01261508, 10151404, 10271408, 12161403, 12161404, 09021404, 01261509, 10271407, 01261507, 10071415, 05011403, 06241407, 08071402, 10091305, 03041403, 01271412, 07281401, and 11061303. Catalog number 905103, 01131542, 11241408, 12161405, 12161406, 0211501, 01211502, 11241406, 02031403, 06241408, 07161407, 09101402, 09281401, 10251303, 05301402, 11241407, and 12101305.
510(k) numbers
["K143616"]
PMA numbers
not on file
Product code
MAX
Firm FEI
3004024955
Firm city
Allendale
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register