FDA Medical Device Recalls (openFDA)

Arrow International Inc: Arrow Continuous Nerve Block Needle, Latex-Free* Rx only*Sterile*Arrow International Inc., 2400 Bernville, Road, Reading, PA 19605 USA

Recall number
Z-0852-05
Recalling firm
Arrow International Inc
Product
Arrow Continuous Nerve Block Needle, Latex-Free* Rx only*Sterile*Arrow International Inc., 2400 Bernville, Road, Reading, PA 19605 USA
Status
Terminated
Initiated
2005-02-17
Posted
2005-06-01
Terminated
2006-05-05
Reason
Possible difficulty in passing the catheter through the needle due to a reduced ID of the needle caused by over crimping of the needle during the manufacturing process.
Root cause
Other
Action
Consignees were notifed by letter on/about 02/17/2005
Quantity
not on file
Distribution
Nationwide, Africa , France, Holland, Rotterdam
Lot, serial or model codes
Product #AB-17040-N, AB-17080-N, AB-17110-N, AB-17140-N, AB-18040-N, AB-18080-N, AB-18110-N, AB-18140-N, AB-19604-K, AB-19604-S, AB-19608-K, AB-19608-S, AB-20604-K, AB-20604-S, AB-20608-K, AB-20608-S
510(k) numbers
["K021567"]
PMA numbers
not on file
Product code
CAZ
Firm FEI
3015859709
Firm city
Reading
Firm state
PA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register