FDA Medical Device Recalls (openFDA)
General Electric Med Systems: GE Single 15 inch TV Monitor Suspension for XT Bridge, Model 46-240485P2, Catalog Number B2056HA
- Recall number
- Z-0851-04
- Recalling firm
- General Electric Med Systems
- Product
- GE Single 15 inch TV Monitor Suspension for XT Bridge, Model 46-240485P2, Catalog Number B2056HA
- Status
- Terminated
- Initiated
- 2004-03-30
- Posted
- 2004-07-20
- Terminated
- 2006-07-24
- Reason
- The ball joint that attaches the single monitor platform to the system boom could fail causing the monitor to fall.
- Root cause
- Other
- Action
- Field modification instructions were issued on 3/30/04 to GE field engineers who are to correct the devices at hospitals. No letter about the correction was sent to hospitals.
- Quantity
- 1236 devices
- Distribution
- The units were distributed nationwide in the United States and worldwide.
- Lot, serial or model codes
- All serial numbers
- 510(k) numbers
- ["K760466","K801320"]
- PMA numbers
- not on file
- Product code
- EAM
- Firm FEI
- 2126677
- Firm city
- Waukesha
- Firm state
- WI
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28