Illumina Inc: Illumina MiSeqDx Cystic Fibrosis Clinical Sequencing Assay in vitro diagnostic system for use on the Illumina MiSeqDx Platform.

Recall number
Z-0850-2015
Recalling firm
Illumina Inc
Product
Illumina MiSeqDx Cystic Fibrosis Clinical Sequencing Assay in vitro diagnostic system for use on the Illumina MiSeqDx Platform.
Status
Terminated
Initiated
2014-11-06
Posted
2014-12-23
Terminated
2015-03-17
Reason
Illumina has confirmed a software limitation with MiSeq Reporter software (version 2.2.31) when using the MiSeqDx Cystic Fibrosis Clinical Sequencing Assay on the MiSeqDx instrument.
Root cause
Under Investigation by firm
Action
A customer notification letter dated 11/12/14 was sent to all customers who purchased the MiSeqDx Cystic Fibrosis Clinical Sequencing Assay. The letter confirms that there is a software limitation with MiSeq Reporter software (version 2.2.31) when using the MiSeqDx Cystic Fibrosis Clinical Sequencing Assay on the MiSeqDx instrument. At certain genomic locations within the region sequenced by the assay, deletions cannot be reported by the software. The letter outlines the potential issues, the results of Illuminia's investigation, and the firm's next steps. Customers' with technical questions or concerns, are instructed to contact their local support team or Illumina Technical Support at techsupport@illumina.com. Customers with questions regarding the field action, are instructed to contact Illumina Regulatory Affairs at RA@illumina.com. Customers will be contacted by 11/13/14 by phone and a phone script was provided that includes the procedure, talking points, actions required, and next steps.
Quantity
19 units
Distribution
US: MD, NY Non-US: France
Lot, serial or model codes
Instrument Serial No. M70106, M00533L, M70114. Lot No. Shipped: K103313, K103461, K103404, K103487.
510(k) numbers
["K132750"]
PMA numbers
not on file
Product code
PFS
Firm FEI
3003218906
Firm city
San Diego
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register