FDA Medical Device Recalls (openFDA)
Kamiya Biomedical Company: K-ASSAY CRP(2) Reagent. Kit is labeled K-ASSAY CRP (2) Latex Particle Enhance ITA. Kit contains 2 vials. One vial labeled K-ASSAY CRP(2) R-1 30 mL 170 mM Glycine Buffer, via has a white cap. The other vial labeled K-ASSAY CRP(2) R-2 20 mL Latex Suspension human CRP rabbit antibodies, vial has a red cap.
- Recall number
- Z-0850-06
- Recalling firm
- Kamiya Biomedical Company
- Product
- K-ASSAY CRP(2) Reagent. Kit is labeled K-ASSAY CRP (2) Latex Particle Enhance ITA. Kit contains 2 vials. One vial labeled K-ASSAY CRP(2) R-1 30 mL 170 mM Glycine Buffer, via has a white cap. The other vial labeled K-ASSAY CRP(2) R-2 20 mL Latex Suspension human CRP rabbit antibodies, vial has a red cap.
- Status
- Terminated
- Initiated
- 2006-04-11
- Posted
- 2006-05-06
- Terminated
- 2006-05-16
- Reason
- Kits may contain 2 vials of the same reagent instead of 1 vial of buffer reagent and 1 vial of antibody reagent
- Root cause
- Other
- Action
- Customers were contacted via telephone on 4/11/06. Customers were asked to confirm the number of kits in their possession and then open the kits to verify which vials they received in their kits. If the kits contained both vials, then the customer was instructed to keep the kit. If the kits contained the wrong vials, then they were requested to return the kits. This was followed by a letter dated 4/17/06. The letter confirmed the telephone call and asked the customer to sign and return the letter to acknoweldge receipt of the letter.
- Quantity
- 185 kits
- Distribution
- Kits were distributed to laboratories nationwide. One foreign account in Singapore.
- Lot, serial or model codes
- Lot No.: 666011 EXP.: 07/2007
- 510(k) numbers
- ["K992312"]
- PMA numbers
- not on file
- Product code
- DCK
- Firm FEI
- 2084025
- Firm city
- Seattle
- Firm state
- WA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28