Skip to the content

FDA Medical Device Recalls (openFDA)

General Electric Med Systems: GE Single 15 inch TV Monitor Suspension with Rail, Model 46-240485P1, Catalog number B0256EH

Recall number
Z-0850-04
Recalling firm
General Electric Med Systems
Product
GE Single 15 inch TV Monitor Suspension with Rail, Model 46-240485P1, Catalog number B0256EH
Status
Terminated
Initiated
2004-03-30
Posted
2004-07-20
Terminated
2006-07-24
Reason
The ball joint that attaches the single monitor platform to the system boom could fail causing the monitor to fall.
Root cause
Other
Action
Field modification instructions were issued on 3/30/04 to GE field engineers who are to correct the devices at hospitals. No letter about the correction was sent to hospitals.
Quantity
1236 devices
Distribution
The units were distributed nationwide in the United States and worldwide.
Lot, serial or model codes
All serial numbers
510(k) numbers
["K760466","K801320"]
PMA numbers
not on file
Product code
EAM
Firm FEI
2126677
Firm city
Waukesha
Firm state
WI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register