FDA Medical Device Recalls (openFDA)
Illumina Inc: Illumina MiSeqDx Universal Kit 1.0, PN 15039608 The Illumina MiSeqDx Universal Kit 1.0 is a set of reagents and consumables used in the processing of human genomic DNA samples derived from peripheral whole blood, and in the subsequent targeted sequencing of the resulting sample libraries. User-supplied analyte specific reagents are required for the preparation of libraries targeting specific genomic regions of interest. The MiSeqDx Universal Kit 1.0 is intended for use with the MiSeqDx instrument.
- Recall number
- Z-0849-2015
- Recalling firm
- Illumina Inc
- Product
- Illumina MiSeqDx Universal Kit 1.0, PN 15039608 The Illumina MiSeqDx Universal Kit 1.0 is a set of reagents and consumables used in the processing of human genomic DNA samples derived from peripheral whole blood, and in the subsequent targeted sequencing of the resulting sample libraries. User-supplied analyte specific reagents are required for the preparation of libraries targeting specific genomic regions of interest. The MiSeqDx Universal Kit 1.0 is intended for use with the MiSeqDx instrument.
- Status
- Terminated
- Initiated
- 2014-11-13
- Posted
- 2014-12-23
- Terminated
- 2015-03-18
- Reason
- It was discovered a software limitation with MiSeq Reporter software (version 2.2.31) when using the MiSeqDx Universal Kit 1.0 on the MiSeqDx instrument. The software cannot report insertions or deletions at specific genomic locations with respect to the primer.
- Root cause
- Software design
- Action
- A customer letter dated November 13, 2014, was sent to all customers who is using the MiSeqDx Universal Kit 1.0 on the MiSeqDx instrument. The letter informs the customers that Illumina has recently confirmed a software limitation with MiSeq Reporter software (version 2.2.31) when using the MiSeqDx Universal Kit 1.0 on the MiSeqDx instrument. The software cannot report insertions or deletions at specific genomic locations with respect to the primer. The customer notification outlines the potential issues, the results of investigation, actions required, and next steps. Customers are instructed to return the attached Verification form to confirm their receipt of the letter and document the results of any patient impact from the limitation. Customers with technical questions or concerns, are instructed to contact their local support team or Illumina Technical Support at techsupport@illumina.com. Customers with questions regarding the field action, are instructed to contact Illumina Regulatory Affairs at RA@illumina.com. For further questions, please call (858) 736-3861.
- Quantity
- 13 units
- Distribution
- US Distribution to the states of AZ, GA, CA, OH, and VA., and one Internationally to France.
- Lot, serial or model codes
- Serial No. M70130, M70189, M70108, M70165, M70166, M70175, M70172, M70103.
- 510(k) numbers
- ["K132750"]
- PMA numbers
- not on file
- Product code
- PFS
- Firm FEI
- 3003218906
- Firm city
- San Diego
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28