Illumina Inc: Illumina MiSeqDx Universal Kit 1.0, PN 15039608 The Illumina MiSeqDx Universal Kit 1.0 is a set of reagents and consumables used in the processing of human genomic DNA samples derived from peripheral whole blood, and in the subsequent targeted sequencing of the resulting sample libraries. User-supplied analyte specific reagents are required for the preparation of libraries targeting specific genomic regions of interest. The MiSeqDx Universal Kit 1.0 is intended for use with the MiSeqDx instrument.

Recall number
Z-0849-2015
Recalling firm
Illumina Inc
Product
Illumina MiSeqDx Universal Kit 1.0, PN 15039608 The Illumina MiSeqDx Universal Kit 1.0 is a set of reagents and consumables used in the processing of human genomic DNA samples derived from peripheral whole blood, and in the subsequent targeted sequencing of the resulting sample libraries. User-supplied analyte specific reagents are required for the preparation of libraries targeting specific genomic regions of interest. The MiSeqDx Universal Kit 1.0 is intended for use with the MiSeqDx instrument.
Status
Terminated
Initiated
2014-11-13
Posted
2014-12-23
Terminated
2015-03-18
Reason
It was discovered a software limitation with MiSeq Reporter software (version 2.2.31) when using the MiSeqDx Universal Kit 1.0 on the MiSeqDx instrument. The software cannot report insertions or deletions at specific genomic locations with respect to the primer.
Root cause
Software design
Action
A customer letter dated November 13, 2014, was sent to all customers who is using the MiSeqDx Universal Kit 1.0 on the MiSeqDx instrument. The letter informs the customers that Illumina has recently confirmed a software limitation with MiSeq Reporter software (version 2.2.31) when using the MiSeqDx Universal Kit 1.0 on the MiSeqDx instrument. The software cannot report insertions or deletions at specific genomic locations with respect to the primer. The customer notification outlines the potential issues, the results of investigation, actions required, and next steps. Customers are instructed to return the attached Verification form to confirm their receipt of the letter and document the results of any patient impact from the limitation. Customers with technical questions or concerns, are instructed to contact their local support team or Illumina Technical Support at techsupport@illumina.com. Customers with questions regarding the field action, are instructed to contact Illumina Regulatory Affairs at RA@illumina.com. For further questions, please call (858) 736-3861.
Quantity
13 units
Distribution
US Distribution to the states of AZ, GA, CA, OH, and VA., and one Internationally to France.
Lot, serial or model codes
Serial No. M70130, M70189, M70108, M70165, M70166, M70175, M70172, M70103.
510(k) numbers
["K132750"]
PMA numbers
not on file
Product code
PFS
Firm FEI
3003218906
Firm city
San Diego
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register