Skip to the content

FDA Medical Device Recalls (openFDA)

USA Instruments Inc.: Prima III TotalSENSE Torso Coil (Body Array Coil), labeled as Manufactured by: USA Instruments, Inc. This device is a component of the Magnetom TRIO 3.0T MRI Systems, manufactured by Siemens AG Medical Solutions.

Recall number
Z-0849-04
Recalling firm
USA Instruments Inc.
Product
Prima III TotalSENSE Torso Coil (Body Array Coil), labeled as Manufactured by: USA Instruments, Inc. This device is a component of the Magnetom TRIO 3.0T MRI Systems, manufactured by Siemens AG Medical Solutions.
Status
Terminated
Initiated
2004-03-19
Posted
2004-07-20
Terminated
2012-03-01
Reason
A defective capacitor in the Body Array Coil may result in unstable, unreliable performance, with excessive heating of patient's body.
Root cause
Other
Action
The firm notified the regional Siemens Service Centers by letter on 3/19/2004. The affected coils will be returned to the recalling firm for modification.
Quantity
79
Distribution
The device was shipped to domestic consignees located CT, NY, PA, NC, GA, FL, OH, IN, MN, IL, UT, CA, OR; and foreign consignees in Austria, Denmark, Spain, Hong Kong, China, Netherlands, Japan, Taiwan, Korea, Australia, Turkey, Brazil, Germany.
Lot, serial or model codes
USAI P/N: 100352, USAI S/Ns: 310 to 388 (Rev 1 and 2); Siemens P/N: 7578789, Siemens S/Ns: 1010 to 1088.
510(k) numbers
["K030371"]
PMA numbers
not on file
Product code
LNH
Firm FEI
3005214683
Firm city
Aurora
Firm state
OH
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register