Skip to the content

FDA Medical Device Recalls (openFDA)

Fresenius Kabi USA, LLC: Ivenix Infusion System (IIS), Large Volume Pump. Model Number: LVP-0004.

Recall number
Z-0848-2026
Recalling firm
Fresenius Kabi USA, LLC
Product
Ivenix Infusion System (IIS), Large Volume Pump. Model Number: LVP-0004.
Status
Open, Classified
Initiated
2025-11-03
Posted
2025-12-01
Terminated
not on file
Reason
Downstream Occlusion alarms may occur during procedures using low flow infusion rates when a check valve* is added to the Ivenix Administration Set.
Root cause
Labeling design
Action
Fresenius Kabi notified consignees on about 11/03/2025 via emailed letter. Consignees were informed of the issue and instructed to account for additional backpressure if check valve is used when configuring the downstream occlusion pressure alarm setting, adjust the occlusion alarm pressure to allow for a more balanced approach between occlusion detection and avoidance of unnecessary alarms, and continue to follow all Instructions for Use and institutional policies regarding infusion pump setup, monitoring, and alarm response. Consignees were requested to ensure all relevant persons within the organization are informed about this letter and the actions as described per the enclosed Tip Sheet on How to Use a Check Valve with the Ivenix Pump & Administration Sets. Additionally, consignees were requested complete and return the Customer Reply Form. Fresenius Kabi plans to update the Instructions for Use to include the information presented in the "Required Actions for Users", "Important Note on Check Valve Usage" and/or Tip Sheet on "How to Use a Check Valve with the Ivenix Pump & Administration Sets" for which scheduling is still being finalized.
Quantity
15,862 units
Distribution
US Nationwide distribution in the sates of CA, CO, FL, GA, ID, IL, MD, MI, MN, MS, NE, NJ, NV, OK, OR, SC, TX, UT, VA, WA, WI.
Lot, serial or model codes
Model Number: LVP-0004. UDI-DI: 00811505030320.
510(k) numbers
["K183311"]
PMA numbers
not on file
Product code
FRN
Firm FEI
3014732157
Firm city
North Andover
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register