Skip to the content

FDA Medical Device Recalls (openFDA)

Maquet Cardiovascular Us Sales, Llc: Compressor Mini To provide a supply of dry, filtered compressed air for a medical respiratory ventilator or anaesthesia machine that meet the specifications of the compressor mini. compressor mini is intended to be operated by healthcare providers, physicians, nurses, and technicians. Compressor mini is to be used only for bedside application within the hospital environment. Compressor mini is neither intended nor suitable for use during in-hospital patient transportation or during ambulance or air transportation. Compressor mini is not suitable for use with MRI.

Recall number
Z-0848-2014
Recalling firm
Maquet Cardiovascular Us Sales, Llc
Product
Compressor Mini To provide a supply of dry, filtered compressed air for a medical respiratory ventilator or anaesthesia machine that meet the specifications of the compressor mini. compressor mini is intended to be operated by healthcare providers, physicians, nurses, and technicians. Compressor mini is to be used only for bedside application within the hospital environment. Compressor mini is neither intended nor suitable for use during in-hospital patient transportation or during ambulance or air transportation. Compressor mini is not suitable for use with MRI.
Status
Terminated
Initiated
2013-12-05
Posted
2014-01-28
Terminated
2015-02-05
Reason
Maquet Compressor Mini may deliver compressed air at a temperature higher than specified.
Root cause
Material/Component Contamination
Action
Maquet sent a product recall urgent-medical device field correction letter/customer fax response via Fed Ex on December 5, 2013, to their customers. Customers are advised of the issues with the device and provided instructions. The correction will include an update with replacement of the affected connector and if needed replacement of the thermoelectric cooler if it damaged. These operations will be performed by an authorized Maquet representative. Any questions can be directed to customer service at 1-888-627-8383 option 2 followed by option 3 Monday through Friday between the hours of 8 am and 6 pm EST. Customers should completed and return the enclosed confirmation form or scan a copy to field actions@maquet.com or fax (973) 807-9290.
Quantity
237 units
Distribution
Worldwide Distribution US, Canada and Japan.
Lot, serial or model codes
Part number 6481779
510(k) numbers
["K023354"]
PMA numbers
not on file
Product code
BTI
Firm FEI
3008355164
Firm city
Wayne
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register