FDA Medical Device Recalls (openFDA)
Maquet Cardiovascular Us Sales, Llc: Compressor Mini To provide a supply of dry, filtered compressed air for a medical respiratory ventilator or anaesthesia machine that meet the specifications of the compressor mini. compressor mini is intended to be operated by healthcare providers, physicians, nurses, and technicians. Compressor mini is to be used only for bedside application within the hospital environment. Compressor mini is neither intended nor suitable for use during in-hospital patient transportation or during ambulance or air transportation. Compressor mini is not suitable for use with MRI.
- Recall number
- Z-0848-2014
- Recalling firm
- Maquet Cardiovascular Us Sales, Llc
- Product
- Compressor Mini To provide a supply of dry, filtered compressed air for a medical respiratory ventilator or anaesthesia machine that meet the specifications of the compressor mini. compressor mini is intended to be operated by healthcare providers, physicians, nurses, and technicians. Compressor mini is to be used only for bedside application within the hospital environment. Compressor mini is neither intended nor suitable for use during in-hospital patient transportation or during ambulance or air transportation. Compressor mini is not suitable for use with MRI.
- Status
- Terminated
- Initiated
- 2013-12-05
- Posted
- 2014-01-28
- Terminated
- 2015-02-05
- Reason
- Maquet Compressor Mini may deliver compressed air at a temperature higher than specified.
- Root cause
- Material/Component Contamination
- Action
- Maquet sent a product recall urgent-medical device field correction letter/customer fax response via Fed Ex on December 5, 2013, to their customers. Customers are advised of the issues with the device and provided instructions. The correction will include an update with replacement of the affected connector and if needed replacement of the thermoelectric cooler if it damaged. These operations will be performed by an authorized Maquet representative. Any questions can be directed to customer service at 1-888-627-8383 option 2 followed by option 3 Monday through Friday between the hours of 8 am and 6 pm EST. Customers should completed and return the enclosed confirmation form or scan a copy to field actions@maquet.com or fax (973) 807-9290.
- Quantity
- 237 units
- Distribution
- Worldwide Distribution US, Canada and Japan.
- Lot, serial or model codes
- Part number 6481779
- 510(k) numbers
- ["K023354"]
- PMA numbers
- not on file
- Product code
- BTI
- Firm FEI
- 3008355164
- Firm city
- Wayne
- Firm state
- NJ
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as