Skip to the content

FDA Medical Device Recalls (openFDA)

bioMerieux, Inc.: VITEK 2 AST-N351 Test Kit

Recall number
Z-0847-2019
Recalling firm
bioMerieux, Inc.
Product
VITEK 2 AST-N351 Test Kit
Status
Terminated
Initiated
2018-12-13
Posted
not on file
Terminated
2021-08-16
Reason
False Positive ESBL Phenotype
Root cause
Software Design Change
Action
On 12/13/2018 bioMerieux submitted a communication to their distributor in the UK stating the following: Our Ref: 4191-1 FSCA Dear Valued bioMrieux Customer, Our records indicate that your laboratory uses the VITEK 2 System in conjunction with the VITEK 2 AST-N351 test kit. Description of Issue: bioMrieux has received reports from VITEK 2 customers that an unexpected ESBL (Extended Spectrum Beta Lactamase) phenotype has been proposed for some Escherichia coli strains in conjunction with the VITEK 2 AST-N351 (ref. 421257) test kit. When an ESBL phenotype is proposed by VITEK 2 Systems Software AES (Advanced Expert System), the result is stopped for review by the user. The intent of the review is that the laboratory personnel take measures to confirm the proposed ESBL phenotype makes sense based on all information available, and perform confirmatory testing if deemed appropriate. Internal bioMrieux investigations concluded that within the VITEK 2 Systems Software version 8.01/9.01 AES (Advanced Expert System) module, the MIC (Minimum Inhibitory Concentration) ranges for Escherichia coli and pipericillin/tazobactam were changed for the Acquired Penicillinase and Inhibitor Resistant Penicillinase phenotypes. These changes consisted of removing MICs >=128 mg/L as possible MICs for both phenotypes. This change revealed an issue in the AES definition of the cefuroxime MIC range for Escherichia coli and ESBLs (the MIC range is too broad; cefuroxime MIC <=8 mg/L should prevent the ESBL phenotype proposal). This, in addition to the specific antibiotic configuration of the AST-N351 card and absence of the ESBL test, is the root cause of the reported increase in ESBL phenotype proposal for the AST-N351 card. Implementation of a bioART" rule with the following criteria will prevent inappropriate proposal of the ESBL phenotype related to this anomaly (reference Appendix A for detailed instructions): " If the card is AST-N351 " And the ph
Quantity
30,292 cartons (20 cards per carton)
Distribution
UK and Ireland
Lot, serial or model codes
Ref: 421257 ALL LOTS
510(k) numbers
not on file
PMA numbers
not on file
Product code
LON
Firm FEI
3002769706
Firm city
Durham
Firm state
NC
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28