Natus Medical Incorporated: Replacement LED PCB Kit, part number 001840, of the Natus neoBLUE (aka neoBLUE 2) Phototherapy system. Neonatal phototherapy.

Recall number
Z-0847-2015
Recalling firm
Natus Medical Incorporated
Product
Replacement LED PCB Kit, part number 001840, of the Natus neoBLUE (aka neoBLUE 2) Phototherapy system. Neonatal phototherapy.
Status
Terminated
Initiated
2014-11-19
Posted
2014-12-22
Terminated
2015-07-20
Reason
Replacement LED Board kit of Natus neoBLUE2 Phototherapy system distributed after February 16, 2012 is subject of a Field Safety Notification because the system will exhibit a higher light intensity output than the original device. While some customers may prefer using these lights at the higher intensity levels, some are concerned that patients may receive more intensity than desired.
Root cause
Other
Action
Natus Medical has begun mailing the Urgent Recall Notification/Field Safety Notice neoBLUE LED Board Replacements letter (DOC-012737B), dated November 5, 2014, to US consignees. Natus Medical also send the Product Advisory neoBLUE LED Board Replacements letter (DOC-0012738), dated October 30, 2014, to International Distribution Partners. Both US and International consignees also received the TECHNICAL BULLETIN LED Board Kit, neoBLUE 2 LED Phototherapy System, p/n: 001840, Treatment Intensity with revised LED board for neoBLUE 2 systems (p/n: 040904 & 040906) [DOC 008353]; and the Label, neoBLUE Intensity/Height (DOC 011561). Consignees are instructed to measure the light intensity of their devices and to document that information of the device using an adhesive label that is provided. The neoBLUE 3 system and its replacement panel sold today are not affected by this issue. If there are any questions regarding this advisory letter or the attached Technical Bulletin, please call Natus Technical Service at 888-496-2887.
Quantity
1907 total (1413 in US and 494 international)
Distribution
worldwide
Lot, serial or model codes
All distributed Natus neoBLUE 2 systems with LED PCB replacements received after February 16, 2012. For use only with neoBLUE devices manufactured between 2002 and 2007 (neoBLUE2)
510(k) numbers
["K022196"]
PMA numbers
not on file
Product code
LBI
Firm FEI
3018859
Firm city
Seattle
Firm state
WA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register