Skip to the content

FDA Medical Device Recalls (openFDA)

Linvatec Corp. dba ConMed Linvatec: The Power Cord Reel is labeled as follows: REF, VP6547, QTY1, POWER CORD REEL, Rx ONLY. These Video Carts are intended for storage of Endoscopic equipment that is used in single or multiple surgical specialties MANUFACTURED FOR: Linvatec, LINVATEC CORPORATION, LARGO, FL 33773 USA.

Recall number
Z-0847-2011
Recalling firm
Linvatec Corp. dba ConMed Linvatec
Product
The Power Cord Reel is labeled as follows: REF, VP6547, QTY1, POWER CORD REEL, Rx ONLY. These Video Carts are intended for storage of Endoscopic equipment that is used in single or multiple surgical specialties MANUFACTURED FOR: Linvatec, LINVATEC CORPORATION, LARGO, FL 33773 USA.
Status
Terminated
Initiated
2010-02-23
Posted
2011-01-04
Terminated
2012-03-22
Reason
ConMed Linvatec, Endoscopy Division, dba. ConMed Linvatec, Largo, Florida is recalling five products: (1.) Video Cart Premium 120 VAC (VP6501), (2.) Video Cart Multi-Specialty 120 VAC (VP6502), (3.) Video Cart Articulating Arm 120V (VP6504), (4.) Video Cart (VP8500), and (5.) Power Cord Assembly Self Retracting (VP6547). These products are being recalled due to there is a possibility that the Vid
Root cause
Component design/selection
Action
ConMed Linvatec sent URGENT MEDICAL DEVICE CORRECTION NOTIFICATION letters to their customers by FedEx. The letter identified the product, the problem, and the action to be taken by the customer. Customers were asked to review their plugs for damage and determine if they were affected. Distributors were asked to contact affected customers and schedule delivery of their replacement plugs within 30 days. If the account no longer has the product in their possession, or if they have transferred the product, they were asked to respond as such via a Reply form. Customers were asked to document delivery of replacement plugs on the Sales Rep Reply Form and mail back or return by fax to 1-888-370-3062. For questions regarding this recall call the Customer Service Department at 1-888-292-0100 or e-mail Custserv1@Invatec.com.
Quantity
77 ea
Distribution
Worldwide Distribution - USA including AK, AL, AR, AZ, CA, CO, CT, FL, GA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MT, NC, NE, NH, NJ, NM, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, VT, WA, WI, WV, AND WY and the countries of Canada, Chile, Colombia, Denmark, Korea, Mexico, Singapore, South Africa, United Arab Emirates, and Venezuela.
Lot, serial or model codes
Power Cord Assembly Self Retracting (Catalog Number: VP6547)  Lot Number: 1160 1533 1590 1959 2534 2558 2595 2724 2798 2905 3009 3269 3428 3831 3831 4099 4498 4894 5209
510(k) numbers
not on file
PMA numbers
not on file
Product code
KQM
Firm FEI
1017294
Firm city
Largo
Firm state
FL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register