FDA Medical Device Recalls (openFDA)
Codman & Shurtleff, Inc.: NEUROPATH 5Fx100 GUIDING CATH HH Product Code: GMC51000000 Intended for use in the coronary, peripheral, and neurovasculature for intravascular introduction of interventional/diagnostic devices
- Recall number
- Z-0846-2013
- Recalling firm
- Codman & Shurtleff, Inc.
- Product
- NEUROPATH 5Fx100 GUIDING CATH HH Product Code: GMC51000000 Intended for use in the coronary, peripheral, and neurovasculature for intravascular introduction of interventional/diagnostic devices
- Status
- Terminated
- Initiated
- 2013-01-11
- Posted
- 2013-02-20
- Terminated
- 2015-08-10
- Reason
- Sterility of device may be compromised due to lack of package integrity
- Root cause
- Component design/selection
- Action
- Codman sent an Urgent Voluntary Product Recall Notification letter dated January 11, 2013, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were advised to check all inventory to determine if they have any affected product, use the included instructions to report inventory status, return and obtain replacement product. Customers were instructed to contact your local Codman Representative or Stericycle at 1-866-201-9077. For questions regarding this recall call 508-880-8100.
- Quantity
- 234 units
- Distribution
- Worldwide Distribution - USA (nationwide) and Internationally to Austria, Belgium, Canada, Chile, Costa, Rica, Czech Republic, Denmark, France, Germany, Great Britian/UK, Ireland, Italy, Lithuania, Malaysia, Netherlands, New Zealand, Norway, Poland, Portugal, Singapore, Slovakia, Spain, and Switzerland.
- Lot, serial or model codes
- All unexpired lots
- 510(k) numbers
- ["K052004"]
- PMA numbers
- not on file
- Product code
- DQY
- Firm FEI
- 3002947440
- Firm city
- Raynham
- Firm state
- MA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28