Skip to the content

FDA Medical Device Recalls (openFDA)

Lin-Zhi International Inc: Barbiturate Enzyme Immunoassay; R1 Antibody/Substrate Reagent (5000 tests and 500 tests); Contains a mixture of monoclonal and polyclonal anti-barbiturate antibody, glucose 6-phosphate (G6P) and nicotinamide adenine dinucleotide (NAD), stabilizers, and with sodium azide as preservative. and Barbiturate Enzyme Immunoassay R2; Enzyme Conjugate Reagent (5000 tests and 500 test packaging); Contains glucose-6-phosphate dehydrogenase (G6PDH) labelled with barbiturate in buffer with sodium azide as preservative. For in-vitro diagnostic use only. Lin-Zhi International Inc. 670 Almanor Ave., Sunnyvale, CA 94085 Product Usage: Drug testing

Recall number
Z-0846-2012
Recalling firm
Lin-Zhi International Inc
Product
Barbiturate Enzyme Immunoassay; R1 Antibody/Substrate Reagent (5000 tests and 500 tests); Contains a mixture of monoclonal and polyclonal anti-barbiturate antibody, glucose 6-phosphate (G6P) and nicotinamide adenine dinucleotide (NAD), stabilizers, and with sodium azide as preservative. and Barbiturate Enzyme Immunoassay R2; Enzyme Conjugate Reagent (5000 tests and 500 test packaging); Contains glucose-6-phosphate dehydrogenase (G6PDH) labelled with barbiturate in buffer with sodium azide as preservative. For in-vitro diagnostic use only. Lin-Zhi International Inc. 670 Almanor Ave., Sunnyvale, CA 94085 Product Usage: Drug testing
Status
Terminated
Initiated
2011-09-08
Posted
2012-01-24
Terminated
2012-01-25
Reason
Catalog #0140 and 0141 Barbiturate EIA small and large kits were not correctly made to specification such that the two antibodies used in production did not have curves that correctly aligned.
Root cause
Nonconforming Material/Component
Action
Lin-Zhi sent a letter via e-mail dated September 8, 2011 to customers with information regarding the recall. The e-mail identified the product description, lot number, description of nonconformity and corrective actions to be taken. Each customer and distributor was then contacted by regulatory affairs personnel by phone on September 9, 2011 or September 12, 2011. Customers and distributors were contacted to determined whether they wanted a replacement of the affected product. Customers that wished for replacements of the affected product was asked to correctly dispose of the product with the lot numbers listed. For questions or concern call 408-732-3856.
Quantity
72 small kits, 12 large kits.
Distribution
Worldwide Distribution - USA (nationwide) including the states of: AL, CA, FL, IN, MA, NC, NH, TX and the countries of: Germany and United Kingdom.
Lot, serial or model codes
Large Kits Catalog number 0141, lot # 1105052, exp 11/25/12; small kits catalog number 0140, lot # 1105053, exp 11/25/12.
510(k) numbers
["K032764"]
PMA numbers
not on file
Product code
DIS
Firm FEI
3003610499
Firm city
Sunnyvale
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register